From: The Scandinavian multicenter hemodynamic evaluation of the SJM Regent aortic valve
Valve size: | 19 | 21 | 23 | 25 | |
---|---|---|---|---|---|
(n): | (4) | (18) | (23) | (16) | |
Mean pressure (mmHg): | p: | ||||
At rest: | 11.4 ± 6.2 | 9.2 ± 6.6 | 7.7 ± 3.3 | 6.0 ± 2.8 | n.s |
Peak stress: | 27.3 ± 22.0 | 21.1 ± 10.0 | 16.9 ± 8.5 | 16.7 ± 5.8 | n.s |
EOA (cm2): | |||||
At rest: | 1.6 ± 0.6 | 1.8 ± 0.4 | 1.8 ± 0.5 | 2.8 ± 1.0 | < 0. 001 |
Peak stress: | 2.2 ± 1.5 | 1.9 ± 0.4 | 1.8 ± 0.5 | 2.7 ± 0.7 | < 0.02 |
EOA index (cm2/m2): | |||||
At rest: | 1.0 ± 0.4 | 1.0 ± 0.2 | 1.0 ± 0.3 | 1.4 ± 0.6 | n.s. |
Peak stress: | 1.3 ± 0.9 | 1.0 ± 0.4 | 1.2 ± 0.4 | 1.3 ± 0.4 | n.s. |