Table 5 Gender comparison of external measurements, aortic lengths and diameters

From: A statistical model that predicts the length from the left subclavian artery to the celiac axis; towards accurate intra aortic balloon sizing

  male (n = 23) female (n = 17)
  mean S.D mean S.D
Height(cm)* 171.09 7.98 160.97 9.20
Weight* 60.30 13.93 51.94 8.51
BSA* 1.70 0.20 1.28 0.14
BMI 20.50 3.93 20.10 3.25
Age 76.57 9.15 68.41 16.24
Measur1* 27.11 1.88 24.97 1.80
Measur2* 47.38 3.03 45.41 2.83
Measur3* 24.83 1.72 23.04 1.76
Measur4 45.10 2.88 43.53 2.75
Measur5* 51.61 3.31 48.92 2.88
Dthorax 26.24 2.53 25.15 3.25
Drenal 24.52 2.67 22.94 2.62
Dbifurc 18.70 2.88 18.38 2.06
LSA-CA* 24.41 2.14 22.68 1.84
LSA-Rena* 27.99 2.11 25.81 2.12
LSA-Bifur* 38.13 3.00 34.75 2.95
LSA-Inser* 53.57 3.43 50.84 2.71
Depth 2.01 0.49 2.12 0.37
  1. *p < 0.05 (t-test)