Skip to main content

Table 1 Comparisons of the modified CPB system with the CCPB system

From: Modified cardiopulmonary bypass with low priming volume for blood conservation in cardiac valve replacement surgery

Parameters

CCPB system

Modified CPB system

CPB tubing

Main pump tubing: 210 cm

Main pump tubing: 80 cm

Venous tubing: 135 cm

Venous tubing: 110 cm

Arterial tubing: 120 cm

Arterial tubing: 100 cm

Tubing from the outlet of the oxygenator artery to the arterial line filter: 75 cm

Total length: 290 cm

Total length: 540 cm

Tubing diameter: approximately 10 mm (3/8 inch)

Tubing diameter: approximately 10 mm (3/8 inch)

Priming volume: 228 mL

Priming volume: 424 mL

Oxygenator

Common oxygenator (RX25 type, Terumo)

Oxygenator integrating arterial line filter (FX25 type, Terumo)

Position: 10 cm from the ground

Position: 80 cm from the ground

Priming volume: 250 mL

Priming volume: 260 mL

Arterial line filter

Separate type

Integrated type

Priming volume: 180 mL

Ultrafiltration circuit

Priming volume: 70 mL

Priming volume: 70 mL

Venous reservoir

Priming volume: 576 mL

Priming volume: 192 mL

Assist device

VAVD device

Assist operation

Retrograde autologous priming method

Yes (discharging 100 mL of priming solution)

CO2 flush procedure

Yes

Cardioplegia perfusion

Single HTK solutions perfusion

Single HTK solutions perfusion

Total priming volume

1500 mL

650 mL