Author | Report type | No. of pediatric patients | Age (years) | Stent | Complications | Treatment of complication | Survived? | Follow-up |
---|---|---|---|---|---|---|---|---|
Rousseau (1999) [15] | Case series | 1 | 14 | Not mentioned | None | – | Yes | 11.6 months |
Karmy-Jones (2002) [16] | Case series | 3 | 12 | PTFE/Palmaz | None | – | No | 1 year |
12 | AneuRX 22-mm cuff | None | – | Yes | ||||
14 | Impra/Wallstent | None | – | No | ||||
Wellons (2004) [17] | Case series | 3 | 15 | AneuRx 28- and 26-mm cuffs | None | – | No | Not mentioned |
16 | Gore 23.5-mm cuff × 2 | None | – | No | ||||
17 | AneuRx 20-mm cuff | Endoleak | Repaired with additional AneuRx 20 mm cuff stent | Yes | ||||
Milas (2006) [18] | Case series | 2 | 16 | 23 mm × 2 endografts | None | – | Yes | 13 months |
17 | Medtronic 24 and 22 mm × 2 AneuRx | None | – | Yes | 22 months | |||
Saad (2006) [19] | Case report | 1 | 12 | Zenith 16 × 55 mm and 18 × 55 mm endografts | None | – | Yes | 12 months |
Azizzadeh (2009) [20] | Case report | 1 | 14 | Excluder (Gore) | None | – | Yes | Not mentioned |
Gunabushanam (2009) [21] | Case report | 1 | 11 | AneuRx 16 mm | None | – | Yes | 10 months |
Goldstein (2012) [22] | Case series | 4 | 14 | 2 Cheatham Platinum stent (20 × 39, 20 × 28 mm) | Stent fracture with no loss of integrity | – | Yes | 24 months |
13 | 2 Cheatham Platinum stent (20 × 45, 20 × 34 mm) | None | – | Yes | ||||
14 | Cheatham Platinum stent (18 × 39 mm) | Recurrent Aneurysm at 24 month follow up | Repaired with second Cheatham Platinum stent (18 × 39 mm) | Yes | ||||
11 | Cheatham Platinum stent (16 × 39 mm) | None | Yes | |||||
Hosn (2016) [23] | Case report | 1 | 8 | zenith alpha 18 mm x 105 mm | None | – | Yes | |
Gombert (2016) [24] | Case report | 1 | 15 | stent prosthesis (ZTEG-2P- 24-115-PF) | None | Yes | 6 months | |
Nicolas (2020) [25] | Case series | 2 | 7 | Cheatham Platinum (16 × 39 mm) | No | Not mentioned | ||
12 | Talent Iliac limb stent graft (18 mm) | Endoleak | Repaired with 15 mm Palmaz Balloon mounted stent inflated to 18 mm | Yes | Not mentioned | |||
Menini stahlschmidt (2010) [26] | Case report | 1 | 11 | – | – | – | Yes | 6 months |
Deeny S (2015) [27] | Case report | 1 | 16 months | iCast stent grafts 16, 22 and 22 mm in overlapping fashion | No | – | Yes | 6 months |
Stringel (2018) [28] | Case Report | 1 | 11 | Cheetham platinum (14 × 45 mm) | Endoleak | Repaired with two Cheatham platinum stetns34 mm and 28 mm | Yes | 20 months |
Hasjim B.J (2019) [14]* | Retrospective review | 124 | 16 (Median) | Multiple (summary in text) | 4.5 years (mean) |