Table 1 Summary of findings of published studies comparing plasma free hemoglobin (PFHb) and haptoglobin between centrifugal pumps and roller pumps

From: Hemolysis and cardiopulmonary bypass: meta-analysis and systematic review of contributing factors

Pump, centrifugal compared to roller in cardiac surgery patients underwent CPB

Outcomes

№ of participants (studies) Follow-up

Certainty of the evidence (GRADE)

Relative effect (95% CI)

Anticipated absolute effects

Risk with Roller

Risk difference with Pump, Centrifugal

Plasma Free Hemoglobin

168 (6 RCTs)

Moderate

The mean plasma Free Hemoglobin was 0

MD 22.27 lower

(35.04 lower to 9.5 lower)

Haptoglobin

112 (3 RCTs)

High

The mean haptoglobin was 0

MD 24.81 higher

(13.68 higher to 35.94 higher)

  1. *The risk in the intervention group (and its 95% confidence interval) is based on the assumed risk in the comparison group and the relative effect of the intervention (and its 95% CI)
  2. CI confidence interval, MD mean difference
  3. GRADE Working Group grades of evidence
  4. High certainty: we are very confident that the true effect lies close to that of the estimate of the effect
  5. Moderate certainty: we are moderately confident in the effect estimate: the true effect is likely to be close to the estimate of the effect, but there is a possibility that it is substantially different
  6. Low certainty: our confidence in the effect estimate is limited: the true effect may be substantially different from the estimate of the effect
  7. Very low certainty: we have very little confidence in the effect estimate: the true effect is likely to be substantially different from the estimate of effect