Advanced mechanical circulatory support for post-cardiotomy cardiogenic shock: a 20-year outcome analysis in a non-transplant unit

Abstract

Background

Post-cardiotomy cardiogenic shock (PCCS) has an incidence of 2–6 % after routine adult cardiac surgery. 0.5–1.5 % are refractory to inotropic and intra-aortic balloon pump (IABP) support. Advanced mechanical circulatory support (AMCS) can be used to salvage carefully selected number of such patients. High costs and major complication rates have lead to centralization and limited funding for such devices in the UK. We have looked the outcomes of such devices in a non-transplant, intermediate-size adult cardiothoracic surgery unit.

Methods

Inclusion criteria included any adult patient who had received salvage veno-arterial extra-corporeal membrane oxygenation (V-A ECMO) or a ventricular assist device (VAD) for PCCS refractory to IABP and inotropic support following cardiac surgery from April 1995-April 2015.

Results

Sixteen patients met the inclusion criteria. Age range was 34–83 years (median 71). There was a male predominance of 12 (75 %). Overall, 15 (94 %) had received ECMO of which, 10 (67 %) had received central ECMO and 5 (33 %) had received peripheral ECMO. One patient (6 %) had a VAD. The most common complication was haemorrhage. Stroke, femoral artery pseudo-aneurysm, sepsis and renal failure also occurred. Thirty-day survival was 37.5 %. Survival rate to hospital discharge was 31.2 %. All survivors had NYHA class I-II at 24 months follow-up.

Conclusions

Our survival rate is similar to that reported in several previous studies. However, the use of AMCS for refractory PCCS is associated with serious complications. The survivors in our cohort appear to maintain an acceptable quality of life.

Peer Review reports

Background

Post-cardiotomy cardiogenic shock (PCCS) occurs in 2–6 % of patients undergoing surgical revascularization or valvular surgery [14]. Approximately 0.5–1.5 % of patients are refractory to maximal inotropic and intra-aortic balloon pump (IABP) support [5]. Post-cardiotomy cardiogenic shock occurs in perioperative cardiac surgery in patients with normal preoperative myocardial function as well as those with pre-existing impaired function [6]. Refractory PCCS leads rapidly to multi-organ dysfunction and is nearly always fatal [4, 79] without the use of advanced mechanical circulatory support (AMCS). AMCS devices such as extracorporeal membrane oxygenation (ECMO) and ventricular assist devices (VAD) have been used to salvage patients who develop refractory PCCS. Survival to hospital discharge is variable [13, 5, 1012] though long term survival and reasonable functional outcome can be achieved [11, 1315]. However, these devices are associated with serious complications [1, 2, 11, 1618] and are costly [9, 19, 20]. The UK National Health Service’s (NHS) proposal to centralise AMCS funding to a few, larger cardiothoracic units has been controversial [15]. Some argue that such a move would remove this potentially, life-saving resource from those cardiothoracic surgery departments excluded from AMCS funding [15]. This prompted us to assess the outcome of the use of AMCS in a non-transplant, intermediate-sized unit (annual workload approximately 900 to 1000 major cardiac surgical operations) in Edinburgh, UK.

Methods

We retrospectively assessed our experience with AMCS in refractory PCCS over a 20-year period. These data were collected prospectively by the Royal Infirmary of Edinburgh cardiac surgery database. Our inclusion criteria included any adult patient from April 1995 to April 2015 who had received salvage V-A ECMO or VAD for PCCS refractory to IABP and maximal inotropic support following adult cardiac surgery. We gained information regarding the patients’ follow-up status from accessing the cardiology follow-up clinic letters on the TrakCareR system, which captures in-patient and out-patient clinical data.

The AMCS devices utilised by our unit in duration of the study included: The LevitronixR CentriMag II for ECMO and Medtronic Bio-MedicusR 560 for VAD support.

Results

There were 18 patients who met the inclusion criteria. We excluded two patients who had had AMCS for refractory PCCS. One patient was in the paediatric age group and had been operated on prior to relocation of paediatric cardiac surgery services to the Royal Hospital for Sick Children, Glasgow in year 2000. Another patient was excluded due to the lack of recorded clarity in the database on the type of AMCS support used, any potential complications and the short and the long-term outcome of this individual. In the remaining 16 patients the age range was 34–83 years (Median 71 years).

Among the remaining 16 (89 %) patients (Table. 1), there was a large male predominance of 12 (75 %). Five patients (31.25 %) had undergone re-operative cardiac surgery. One patient (6.25 %) had undergone AMCS following the repair of an ascending aortic transection after a road traffic accident. Overall, 15 patients (94 %) had received a single run of V-A ECMO of which number, 10 (67 %) had received central ECMO and 5 (33 %) had received peripheral ECMO. One patient (6 %) had VAD. The mean duration of AMCS was approximately 5.4 days (Range < 1 day – 33 days). The most common procedure-related complication was major haemorrhage. The incidence of major cerebrovascular accident, peripheral limb ischaemia, femoral artery pseudo-aneurysm, septic shock and renal failure requiring renal replacement therapy was 18.75 %, 12.5 %, 6.25 %, 12.5 % and 18.75 % respectively (Fig. 1). Logistic EuroSCORE ranged from 2.08 to 73.26 (mean 20.21, SD = 17).

Table 1 Illustrates patient characteristics, the type of mechanical circulatory support and outcomes over the 11-year study period
Fig. 1
figure 1

Illustrates the number of complications in our cohort

Most common cause of death (COD) was refractory biventricular failure (37.5 %, Fig. 2), in which group of patients, AMCS was withdrawn. One patient died due to combination of biventricular failure and haemorrhagic shock whilst on VA ECMO. Thirty-day survival was 37.5 % (Fig. 3). Our survival rate to hospital discharge was 31.2 %. Upon reviewing the consultant cardiologist follow-up clinic letters on the TrakCareR database, all survivors who were discharged home were alive at 24 months and had NYHA class I-II functional status on follow-up.

Fig. 2
figure 2

Illustrates the cause of death in our cohort

Fig. 3
figure 3

Kaplan-Meier curve of survival. FU: follow up (months), Cum Survival: cumulative survival

Advanced age, the emergent nature of surgery, pre-existing, preoperative severe left ventricular impairment were identified as possible factors leading to an adverse outcome. These findings however are limited due to the small number of subjects and the retrospective nature of the study.

Discussion & review of the literature

AMCS have been in evolution for almost 50 years. AMCS devices developed in parallel with the cardiopulmonary bypass (CPB) machine by John Gibbon in 1953 [21, 22]. Soon after its inception, the CPB machine was being used to salvage patients following “failed” cardiac surgery. In 1966, Michael DeBakey used the first LVAD to support a patient who was in refractory PCCS [21]. After the first heart transplantation by Christiaan Barnard in 1967 and widespread acceptance of transplantation, AMCS devices have been regularly used to sustain those patients in refractory PCCS to allow time for decision for further management with implantable devices or heart transplantation [21].

There are two forms of ECMO; veno-venous (VV) which is usually utilized to support isolated respiratory failure and veno-arterial (VA) ECMO used to support cardiac, respiratory or mixed cardiorespiratory failure. Both types of ECMO can be applied centrally and peripherally [23, 24]. There are two types of VADs. Short-term, percutaneous VADs, which can be inserted in the catheter laboratory, serve as a temporizing measure to stabilize patients in acute cardiogenic shock. More long-term, surgically-implanted VADs are designed to increase patient’s life span and/or bridge patients to transplantation [15]. VADs can be used to support the failing right ventricle, left ventricle or both. According to the UK VAD registry, scarcity of donor hearts has stimulated consideration of implantable AMCS devices for more prolonged circulatory support [20]. Recent studies have demonstrated that modern continuous flow AMCS devices such as CentriMagR pumps with magnetically levitated rotors and more effective modern oxygenators, can be more easily implanted, leading to much better survival in patients with post-cardiotomy cardiogenic shock in the recent years [2527].

In our literature search we identified 15 major articles. In the largest cohort, Hernandez et al. [3] collated data from 5,735 patients who underwent salvage VAD support for refractory PCCS. They reported 54.1 % survival rate to hospital discharge. They concluded that VAD is a valuable, life-saving therapeutic manoeuvre. Rastan et al. [5] performed ECMO support in 516 patients with refractory PCCS. A total of 24.8 % of patients survived to discharge to the community. However, after 5 years 13.7 % were alive and 17.4 % of patients suffered severe morbidity [5]. In another large cohort of 219 patients Doll et al. [28] reported 24.8 % survival to discharge and 16.9 % 5-year survival for their cohort of refractory PCCS patients who had received salvage ECMO support. Hsu et al [12] reported a cohort of 51 patients who had undergone cardiac surgery and suffered refractory PCCS requiring ECMO circulatory support. The 30-day, 3-month and 1-year mortality rates were 49, 65 and 71 % respectively. Mehta et al. [13] reported a large cohort of 1,279 patients who had undergone VAD for circulatory support for refractory PCCS. They claimed that 584 (45.7 %) patients were successfully weaned from VAD and 323 (25.3 %) patients survived to discharge from the hospital. In a small cohort of 12 patients with refractory PCCS, DeRose et al. [2] reported that 9 patients (75 %) survived to discharge from the hospital with LVAD implantation, 8 (67 %) survived to transplantation and 1 (8 %) successfully underwent explanation of LVAD not requiring transplantation. However, they reported 42 % rate of LVAD related infective complications.

With respect to longer term outcomes; a study by Ko et al. [11] reported a cohort of 76 patients undergoing ECMO support for refractory PCCS. They reported that, although 46 patients (60.5 %) were successfully weaned from ECMO, 20 (26.3 %) survived to discharge. However all survivors were reported to be of New York Heart Association (NYHA) I and II functional status on 32 +/- 22 month follow-up. This quality of life finding is in keeping with the findings in our cohort. Pennington et al. [17] reported refractory PCCS support with VAD with 37 % survival to hospital discharge. They reported that all survivors were “leading active lives”. In 72.7 % of survivors ejection fraction had normalized on follow-up echocardiogram. AMCS devices also have disadvantages. They are currently not funded within the cardiac surgery tariff and are quite costly [9, 19, 20]. According to a study conducted at the University Hospital of South Manchester [9] (an NHS institution), the cost of capital equipment, device maintenance and single-use elements (e.g. tubing) of the CentriMagR device is £3542. The total per patient cost of VAD and ECMO were reported as £15 669 and £8616 respectively. However the costs associated with the device and the ICU stay may vary substantially depending on the duration of AMCS. The cost of implantable VAD per patient receiving a device is £173,841 or US$316,078. The most costly aspect is implantation of the device (£63,830, US$116,056) and the initial hospital stay in the ITU and ward totals as much as £14,500 or US$26,364 [29]. A clinical trial indicated the cost of quality adjusted life year for ECMO in the UK to be £19,252 [30]. Also there are reports [1, 2, 11, 1618] of serious device-related complications including: bleeding, thrombus formation and embolization, cerebrovascular accidents, device infection, limb ischaemia and multi-organ dysfunction syndrome/failure. However, there is evidence that on-the-table, early implantation of AMCS devices, prior to leaving the operating theatre from the initial operation, substantially improves survival as compared to late implantation out-with the operating theatre [6, 26]. Given the reported survival rates and the extent of resources involved, we recommend that each case should be assessed in its own individual merit. Involvement of a cardiac surgeon, other than the surgeon involved in the initial operation, the clinical director of the department of cardiac surgery, as well as involving the perfusionist and the on-call anaesthetist in the decision making process is imperative in gaining optimal outcomes for patients undergoing AMCS for refractory PCCS while optimising the cost-benefit equation.

Conclusion

AMCS devices can be used to salvage some patients with refractory PCCS who would otherwise have not survived. However, ACMS are associated with high rates of severe, systemic and device-related complications as well being costly. We recommend team approach to decision-making and early application of AMCS to the few carefully selected patients with refractory PCCS in order to optimise the cost-benefit equation. Our results reflect findings from previous studies. Our study showed survivors enjoyed a reasonable quality of life.

References

  1. Muehrcke DD, McCarthy PM, Stewart RW, Foster RC, Ogella DA, Borsh JA, et al. Extracorporeal membrane oxygenation for postcardiotomy cardiogenic shock. Ann Thorac Surg. 1996;61:684–91.

    Article  CAS  PubMed  Google Scholar 

  2. DeRose JJ, Umana JP, Argenziano M, Catanese KA, Levin HR, Sun BC, et al. Improved results for postcardiotomy cardiogenic shock with the use of implantable left ventricular assist devices. Ann Thorac Surg. 1997;64:1757–63.

    Article  PubMed  Google Scholar 

  3. Hernandez AF, Grab JD, Gammie JS, O’Brien SM, Hammill BG, Rogers JG, et al. A decade of short-term outcomes in post–cardiac surgery ventricular assist device implantation. Circulation. 2007;116:606–12.

    Article  PubMed  Google Scholar 

  4. Mohite PN, Sabashnikov A, Patil NP, Sáez DG, Zych B, Popov AF, et al. Short-term ventricular assist device in post-cardiotomy cardiogenic shock: factors influencing survival. J Artif Organs. 2014;17(3):228–35.

    Article  PubMed  Google Scholar 

  5. Rastan AJ, Dege A, Mohr M, Doll N, Falk V, Walther T, et al. Early and late outcomes of 517 consecutive adult patients treated with extracorporeal membrane oxygenation for refractory postcardiotomy cardiogenic shock. J Thorac Cardiovasc Surg. 2010;139:302–11.

    Article  PubMed  Google Scholar 

  6. Delgado DH, Rao V, Ross HJ, Verma S, Smedira NG. Mechanical circulatory assistance: state of art. Circulation. 2002;106:2046–50.

    Article  PubMed  Google Scholar 

  7. Beurtheret S, Mordant P, Paoletti X, Marijon E, Celermajer DS, Le’ger P, et al. Emergency circulatory support in refractory cardiogenic shock patients in remote institutions: a pilot study (the cardiac-RESCUE program). Eur Heart J. 2012;34(2):112–20.

    Article  PubMed  Google Scholar 

  8. Goldstein DJ MCO. Mechanical support for postcardiotomy cardiogenic shock. Semin Thorac Cardiovasc Surg. 2000;12(3):220–8.

    Article  PubMed  Google Scholar 

  9. Borisenko O, Wylie G, Payne J, Bjessmo S, Smith J, Firmin R, et al. The cost impact of short-term ventricular assist devices and extracorporeal life support systems therapies on the National Health Service in the UK. Interact Cardiovasc Thorac Surg. 2014;19:41–8.

    Article  PubMed  Google Scholar 

  10. De Robertis F, Birks EJ, Rogers P, Dreyfus G, Pepper JR, Khaghani A. Clinical performance with the levitronix centrimag short-term ventricular assist device. J Heart Lung Transplant. 2006;25:181–6.

    Article  PubMed  Google Scholar 

  11. Ko WJ, Lin CY, Chen RJ, Wang SS, Lin FY, Chen YS, et al. Extracorporeal membrane oxygenation support for adult postcardiotomy cardiogenic shock. Ann Thorac Surg. 2002;73:538–45.

    Article  PubMed  Google Scholar 

  12. Hsu PS, Chen JL, Hong GJ, Tsai YT, Lin CY, Lee CY, et al. Extracorporeal membrane oxygenation for refractory cardiogenic shock after cardiac surgery: predictors of early mortality and outcome from 51 adult patients. Eur J Cardiothorac Surg. 2010;37:328–33.

    PubMed  Google Scholar 

  13. Mehta SM, Aufiero TX, Pae WE, Miller CA, Pierce WS. Results of mechanical ventricular assistance for the treatment of post cardiotomy cardiogenic shock. ASAIO J. 1996;42:211–8.

    CAS  PubMed  Google Scholar 

  14. Pae WE, Pierce WS, Pennock JL, Campbell DB, JA W. Long-term results of ventricular assist pumping in postcardiotomy cardiogenic shock. J Thorac Cardiovasc Surg. 1987;93(3):434–41.

    PubMed  Google Scholar 

  15. Westaby S, Taggart D. Inappropriate restrictions on life saving technology. Heart. 2012;98(15):1117–9.

    Article  PubMed Central  PubMed  Google Scholar 

  16. Xiao XJ, Luo ZX, Ye CX, Fan RX, Yi DH, Ji SY, et al. The short-term pulsatile ventricular assist device for postcardiotomy cardiogenic shock: a clinical trial in China. Artif Organs. 2009;33(4):373–7.

    Article  PubMed  Google Scholar 

  17. Pennington DG, McBride LR, Swartz M, Kanter KR, Kaiser GC, Barner HB, et al. Use of the pierce-donachv ventricular assist device in patients with cardiogenic shock after cardiac operations. Ann Thorac Surg. 1989;47:130–5.

    Article  CAS  PubMed  Google Scholar 

  18. Zumbro GL, Kitchens WR, Shearer G, Harville G, Bailey L, Galloway RF, et al. Mechanical assistance for cardiogenic shock following cardiac surgery, myocardial infarction, and cardiac transplantation. Ann Thorac Surg. 1987;44(1):11–3.

    Article  CAS  PubMed  Google Scholar 

  19. Miller LW, Guglin M, Rogers J. Cost of ventricular assist devices: can we afford the progress? Circulation. 2013;127:743–8.

    Article  PubMed  Google Scholar 

  20. Emin A, Rogers CA, Parameshwar J, MacGowan G, Taylor R, Yonan N, et al. Trends in long-termmechanical circulatory support for advanced heart failure in the UK. Eur J Heart Fail. 2013;15:1185–93.

    Article  PubMed  Google Scholar 

  21. Stewart GC, Giverts MM. Mechanical circulatory support for advanced heart failure: Patients and technology in evolution. Circulation. 2012;125:1304–15.

    Article  PubMed  Google Scholar 

  22. Fou AA. John H. Gibbon. The first 20 years of the heart-lung machine. Tex Heart Inst J. 1997;24(1):1–8.

    PubMed Central  CAS  PubMed  Google Scholar 

  23. Aubron C, Cheng AC, Pilcher D, Leong T, Magrin G, Cooper DJ, et al. Factors associated with outcomes of patients on extracorporeal membrane oxygenation support: a 5-year cohort study. Crit Care. 2013;17(2):1–12.

    Article  Google Scholar 

  24. Interventional procedure overview of extracorporeal membrane oxygenation (ECMO) for acute heart failure in adults. National Institute for Health and Care Excellence (NICE); 2013. p. 1-40. https://www.nice.org.uk/guidance/ipg482/resources/extracorporeal-membrane-oxygenation-ecmo-for-acuteheart-failure-in-adults-1899869988516037.

  25. Mikus E, Tripodi A, Calvi S, Giglio MD, Cavallucci A, Lamarra M. CentriMag venoarterial extracorporeal membrane oxygenation support as treatment for patients with refractory postcardiotomy cardiogenic shock. ASAIO J. 2013;59(1):18–23.

    Article  CAS  PubMed  Google Scholar 

  26. Akay MH, Gregoric ID, Radovancevic R, Cohn WE, Frazier OH. Timely use of centrimag heart assist device improves survival in postcardiotomy cardiogenic shock. J Card Surg. 2011;26(5):548–52.

    Article  PubMed  Google Scholar 

  27. Peura JL, Colvin-Adams M, Francis GS, Grady KL, Hoffman TM, Jessup M, et al. Recommendations for the use of mechanical circulatory support: device strategies and patient selection. Circulation. 2012;126:2648–67.

    Article  PubMed  Google Scholar 

  28. Doll NKB, Borger M, Bucerius J, Kramer K, Schmitt D, et al. Five year results of 219 consecutive patients treated with extracorporeal membrane oxygenation after refractory postoperative cardiogenic shock. Ann Thorac Surg. 2004;77:151–7.

    Article  PubMed  Google Scholar 

  29. Sharples LD, Dyer M, Cafferty F, Demiris N, Freeman C, Banner NR, et al. Cost-effectiveness of ventricular assist device use in the United Kingdom: results from the Evaluation of Ventricular Assist Device Programme in the UK (EVAD-UK). J Heart Lung Transplant. 2006;25(11):1336–43.

    Article  PubMed  Google Scholar 

  30. Stewart NI, Gunning K, Cuthbertson BH. Regionalisation of intensive care and extra-corporeal membrane oxygenation services in the UK: beliefs about the evidence, benefits and harm. J Intensive Care Society. 2012;13(3):244–50.

    Article  Google Scholar 

Download references

Acknowledgements

International Meeting Presentation: Oral presentation & shortlisted for a prize at the World Society of Cardiothoracic Surgery (WSCTS), 2015.

Funding

No funding was claimed for this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Maziar Khorsandi.

Additional information

Competing interest

The authors declare that they have no competing interests.

Authors’ contribution

MK: Primary investigator, presented at the WSCTS 2015, manuscript preparation. KS: Manuscript drafting and preparation. SP: Manuscript editing. RP: Manuscript editing. WW: Manuscript editing. GB: Manuscript editing. VZ: Project supervisor and senior author. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Khorsandi, M., Shaikhrezai, K., Prasad, S. et al. Advanced mechanical circulatory support for post-cardiotomy cardiogenic shock: a 20-year outcome analysis in a non-transplant unit. J Cardiothorac Surg 11, 29 (2016). https://doi.org/10.1186/s13019-016-0430-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-016-0430-2

Keywords