Computer-based comparison of different methods for selecting mitral annuloplasty ring size

Abstract

Background

Ring sizing for mitral valve annuloplasty is conventionally done intraoperatively using specific ‘sizer’ instruments, which are placed onto the valve tissue. This approach is barely reproducible since different sizing strategies have been established among surgeons. The goal of this study is to virtually apply different sizing methods on the basis of pre-repair echocardiography to find out basic differences between sizing strategies.

Methods

In three-dimensional echocardiographs of 43 patients, the mitral annulus and the contour of the anterior mitral leaflet were segmented using MITK Mitralyzer software. Similarly, three-dimensional virtual models of Carpentier-Edwards Physio II annuloplasty rings and their corresponding sizers were interactively generated from computer tomography images. For each patient, the matching annuloplasty ring was selected repeatedly according to popular sizing strategies, such as the height of anterior mitral leaflet, the intercommissural distance and the surface area of anterior mitral leaflet. The areas of the selected rings were considered as the neo-surface area of the mitral annulus after implantation.

Results

The sizing of the mitral valve according to the height of anterior mitral leaflet (mean ring size = 29.9 ± 3.90), intercommissural distance (mean ring size = 37.5 ± 1.92) or surface area of anterior mitral leaflet (mean ring size = 32.7 ± 3.3) led to significantly different measurements (p ≤ 0.01). In contrary to intercommissural distance, height and surface area of the anterior mitral leaflet exhibited significant variations between the patients (p ≤ 0.01). The sizing according to the height of anterior mitral leaflet led to the maximal reduction of the mitral annulus surface area followed by the sizing according to the surface area of anterior mitral leaflet and finally by the intercommissural distance.

Conclusions

This novel comprehensive computer-based analysis reveals that the surveyed sizing methods led to the selection of significantly different annuloplasty rings and therefore underscore the ambiguity of routinely applied annuloplasty sizing strategies.

Peer Review reports

Background

Mitral valve repair is a common procedure in the field of cardiac surgery. Despite the huge experience in this field, there is no standard approach to repair the mitral valve. Most techniques were developed empirically including the implantation of annuloplasty rings in order to reshape, stabilize or downsize the valvular annulus such that a sufficient amount of leaflet tissue is involved in coaptation. As the usage of prosthetic implants of suboptimal size or repair without ring insertion has been shown to adversely affect surgical results [1, 2], choosing the adequate ring size can be seen as critical [3, 4].

To date, one of the most commonly used prosthetic ring is the commercially available Carpentier-Edwards Physio II mitral ring (CE ring), which comes in 9 different sizes (size 28–40) [3]. Over the past decades, different strategies to find out the appropriate ring for various pathologies affecting the mitral valve apparatus have been developed. For choosing the suitable annuloplasty ring, template like translucent “sizer” instruments are used (see Fig. 1b), which are roughly similar in size like the prosthetic ring. They are placed onto the valve in-situ by the surgeon according to specific anatomical relations, such as the distance of the trigonal area, the intercommissural distance, the size of the annulus, the height or the surface area of the anterior mitral leaflet (AML). The AML has been preferably considered for sizing, since its tissue contributes mostly to mitral orifice occlusion upon systole; therefore the annulus should have approximately the size of the AML after ring implantation to allow the remaining AML tissue to coapt. Studies have proposed a vast amount of techniques to use the mentioned landmarks to find the best suitable ring according to the different pathologies and anatomical variations [3, 521]. To complicate the matter, concepts of “truesizing” and “downsizing” have emerged, which means that either the best matching ring is selected for the patient or one or two sizes smaller to reduce the size of the potentially dilated mitral orifice.

Fig. 1
figure 1

a Screenshot of the MITK Mitralyzer plugin with a modelled mitral annulus and anterior mitral leaflet. Furthermore, three anatomical measurements are shown, which are important for ring sizing. b Corresponding measurements on the sizer instrument

The advent of real-time three-dimensional transesophageal echocardiography (3D TEE) has allowed impressive volume renderings of the mitral valve anatomy and dynamics, especially the annulus, the leaflets and the papillary muscles, thereby significantly improving diagnostic capabilities and therapy planning. Additionally, the valve can be segmented from these images such that patient-specific virtual models are created. A specific software plugin called “Mitralyzer” of the freely available Medical Imaging Interaction Toolkit (MITK), which is able to process medical images [22, 23], was developed by our group for interactive and semi-automatic 3dimensional mitral annulus modeling [24]. The software allows for precise virtual reconstruction of the mitral annulus curve and the AML contour using 16 defined points on mitral annulus and the anterior mitral leaflet (see Fig. 1). The MITK plugin is planned to be released as a freely available software package soon.

Our goal was to study the effect of different sizing methods commonly used to implant a CE ring with regard to ring size selection, i.e. to find out if various strategies lead to the same results.

Methods

Intraoperative real-time three-dimensional transesophageal echocardiographs of 43 patients, who underwent mitral valve surgery due to insufficiency, were extracted using Live-3D mode of the Philips ultrasound system iE33 xMatrix and a Philips X7-2 t matrix array transducer (Philips Healthcare, Andover, MA, USA). The images were taken intraoperatively on the beating heart and before initiation of the cardiopulmonary bypass. Only patients with good quality images that involve the whole mitral annulus and the mitral leaflets with the subvalvular apparatus were involved in this study. The images were transferred to a workstation running Philips QLab Software and exported in DICOM Cartesian file format to enable further processing. Using MITK Mitralyzer software with mitral valve images at early systolic phase of the cardiac cycle allowed the interactive generation of mitral annulus models and the contour models of the AML. Carpentier-Edwards Physio II annuloplasty ring sizers and Carpentier-Edwards Physio II annuloplasty rings of size 26–38 were digitized using computer tomography (CT, Somatom Denition, Siemens Medical Systems). The images were exported to DICOM format and a three-dimensional model representing each planar sizer (sizer model) and its corresponding annuloplasty ring (ring model) were made using MITK Mitralyzer. By the help of the software, automatic quantifications could be derived from the models, such as surface area of AML model, height of the AML model, intercommissural distance (same for both models) for each patient (see Fig. 1). From the sizer models the corresponding measurements (width, height, area) were computed automatically. Also the orifice area of each ring model was calculated using delaunay triangulation.

For each patient’s mitral annulus and AML model, we virtually selected the matching sizer model according to the following five different sizing methods: 1) corresponding height of AML (true sizing); 2) corresponding intercommissural distance (true sizing); 3) intercommissural distance with one size smaller (downsizing); 4) intercommissural distance with two sizes smaller (downsizing) and finally 5) corresponding AML surface area (true sizing). The anatomical references for the “true sizing” strategies are illustrated in Fig. 1.

Subsequently, the conformity between the results of the five strategies, i.e. the selected ring size, was compared to each other for each patient separately. The surface area of the chosen rings was considered as the surface area of the corresponding mitral annuli after annuloplasty.

The collected data was analyzed using analysis of variance (ANOVA) and t-test. Results are presented as mean ± standard deviation in mm for measurements and without standard unit for ring sizes. A p-value less than 0.05 was considered significant.

Results

The mean age of the 43 patients at the time of the operation was 63.7 ± 11.8 years. About 77% of the patients were male. Thirty-four patients received mitral valve repair (81%) using Physio II annuloplasty ring. Eight patients (24%) of them needed an additional implantation of artificial polytetrafluoroethylene neochordae during mitral valve repair while sixteen patients (47%) received leaflet repair procedures and seven patients (21%) received both neochordae and leaflet repair. Nine patients underwent mitral valve replacement. More detailed characteristics of the patient cohort are summarized in Table 1.

Table 1 Patient characteristics

Considering the whole patient collective, the sizing of the mitral valve led to significantly different average measurements (p ≤ 0.01) according to the five applied strategies: height of AML (mean = 29.9 ± 3.90), intercommissural distance with true sizing (mean = 37.5 ± 1.92), intercommissural distance with one size smaller (mean = 35.5 ± 1.65) or intercommissural distance with two sizes smaller (mean = 33.6 ± 1.38) and AML surface area sizing (mean = 32.7 ± 3.29). Fig. 2 illustrates this by showing the averaged mitral annulus and AML morphological model for the 43 patients together with the outcomes of the mean “true sizing” strategies (rounded to commercially available sizes).

Fig. 2
figure 2

Averaged mitral annulus and AML model of all 43 patients (white) with the matching annuloplasty rings (green) according to the “true sizing” strategies. Mean ring size values were rounded to real commercially available sizes

Measurements of mitral valve apparatus landmarks are provided in Table 2. They showed significant variations in our patients’ cohort with exception of intercommissural distances. Height of AML measurements in our patients cohort (height AML mean =23.9 ± 3.74 mm) shows a significant variation between the patients (p ≤ 0.01). Intercommissural distance measurements in our patients cohort (mean = 38. 9 ± 6. 15 mm) shows no significant variation (p = 0. 31). The mean AML surface area was 711.13 ± 175.56 mm2 with significant variation between the patients (p ≤ 0. 01).

Table 2 Measurements of the studied patient cohort of 43 patients in systole

The most sizing methods typically rely on measurements of the AML due to its tissue surface contribution to mitral valve closure, such that we did not consider the posterior mitral leaflet. Therefore, we compared the complete surface area of virtually implanted annuloplasty ring with the surface area of the AML. If the implanted ring was smaller than the AML area, we assumed that the implantation of such a ring will lead to a reduction of the visible AML surface area that participates in the mitral valve orifice during systole. The mean reduction in AML surface area when using height of AML for sizing was 116.7 ± 137.05 mm2 (16.4% reduction in AML surface area) while the reduction in AML when using AML surface area sizing was 19.43 ± 132.37 mm2 (1.2% reduction in AML surface area).

Using the intercommissural distance with true sizing (strategy 2), intercommissural distance with one size smaller (strategy 3), or intercommissural distance with two sizes smaller (strategy 4) led to no reduction in AML surface area as the implanted rings were 241.1 ± 164.86 mm2, 125.46 ± 163.06 mm2 and 30.24 ± 163.93 mm2 bigger than the AML surface area respectively.

Mean mitral annulus area was 1413.09 ± 338.31 mm2. The reduction in annulus area when using height of AML for sizing was 818.665 ± 296.3 mm2 (strategy 1) while the reduction in annulus area when using AML area sizing (strategy 5), intercommissural distance with true sizing (strategy 2), intercommissural distance with one size smaller (strategy 3) and intercommissural distance with two sizes smaller (strategy 4) were 721.39 ± 249.45 mm2, 460.818 ± 311.75 mm2, 576.493 ± 315.29 mm2 and 671.715 ± 318.64 mm2 respectively.

The sizing results according to intercommissural distance (strategy 2) and AML surface area (strategy 5) led to the same sizing results in 14 patients (33%). For AML height (strategy 1) und AML surface area (strategy 5), ring sizing was equal in 10 patients (23%). In 4 patients (9%), the sizing results were equal according to AML height (strategy 1) and intercommissural distance (strategy 2). Note that in only 2 of the 43 patients (5%), the ring sizing in all methods was equal (see Fig. 3).

Fig. 3
figure 3

Sizing results for each patient according to the true sizing strategies: intercommissural distance, AML height, and AML surface area

The actually used intraoperative method to size the mitral valve as conducted by surgeons are not documented. However, the mean size of the used mitral annuloplasty rings was 32.11 ± 2.44 and no obvious relation was found between the implanted rings and the virtually proposed rings. From our results, the outcome of AML surface area sizing method (mean = 32.7 ± 3.29) has its closest relationships to the actual implanted ring size.

Discussion

Mitral valve reconstruction remains one of the most challenging fields in cardiac surgery with continuously changing concepts and new emerging ideas. The purpose of the current computer-based study was to find out if some of the currently used intraoperative methods for sizing the mitral valve annulus, as part of mitral valve repair using the CE physio II annuloplasty ring, result in similar ring size choices. The use of echocardiographic measurements to figure out the suitable annuloplasty ring is gaining more importance especially if three-dimensional measurements are performed [25]. The mostly used dimensions are the height of the AML and the intercommissural distance. Furthermore, the effect of different annuloplasty rings on the mitral valve geometry has been studied in-silico [26, 27] and in animal models employing radiopaque markers [28].

In our study, the use of the height of the AML (strategy 1) as reference to implant the annuloplasty ring was associated with the highest reduction in mitral valve orifice area and reduces the visible AML surface occupation area of the orifice; this theoretically implies the highest increase in coaptation area between AML and posterior mitral leaflet (PML). Such an increase in the coaptation area may result in a better competence of the mitral valve but may also lead to displacement of the AML causing systolic anterior movement (SAM) and left ventricular outflow tract obstruction with mitral insufficiency especially in the presence of a prominent PML. It may also cause iatrogenic mitral stenosis due to abundant valve tissue in the mitral valve orifice. Such postulations need further studies and finite element-based patient-specific annuloplasty simulations methods might be viable approach for this, such as shown by Schoch et al. [26]. However, it is crucial to take the height of the into account to avoid such complications. Many surgeon use AML height to better assess the dimensions of the mitral valve dimensions, but they also take the intercommissural distance in consideration to avoid excessive downsizing [3, 5, 14]. Note that in our study we only considered isolated strategies so far to show their single unaffected outcome.

Using the AML surface area (strategy 5) as a reference for sizing respects the AML and leads to less reduction in the mitral valve opening area which seems to be the best method to preserve the AML function but at the same time compromises the size of coaptation area [6], which considered as measure for the valves’ functional integrity. However, the accurate measurement of the AML surface area is more complex than the other chosen parameters, since we only contoured the AML edge but did not model the exact leaflet curvature. When comparing the outcome of AML height and intercommissural distance sizing, AML surface area sizing provides intermediate results for which most of the surgeon are somehow adapted to according to their experience [10, 11, 13].

Our study showed that sizing according to intercommissural distance results in the implantation of bigger annuloplasty rings (see Fig. 3), even if downsizing is used, which would in turn less decrease the opening area of the mitral valve in comparison to the other presented strategies. This implies that the PML need to play a more crucial role in mitral valve competence. Therefore, many surgeon use this dimension as the main sizing strategy while others combine the intercommissural distance with other dimensions to find out the optimal annuloplasty ring for their patients [3, 5, 20].

To further complicate sizing, it has been already shown by Bothe et al. [4], who termed the application of sizer instruments a “voodoo”, that there is a size discrepancy between corresponding rings and sizers which limits the scientific value of this procedure. Our comparison presented in Table 3 after CT imaging and interactive modelling of rings and sizes are in concordance with the findings by Bothe et al. [4].

Table 3 Comparison of measurements from CT-scanned and modelled Physio II annuloplasty ring-sizers and Physio II annuloplasty rings

Note that the valve geometry under physiological loading conditions considerably differs from the cardioplegic valve morphology. Intraoperatively, when sizing is usually conducted, the valve is flaccid and collapsed. The actual difference in size and shape between these two states could be virtually visualized and quantified by our group for the first time [29], using novel accurate optical measurement technology. Future work includes collection of intraoperative valve shape information and the actual conducted sizing procedure to enable improved comparisons. The ultimate goal would be to translate the hitherto ambiguous intraoperative experiences of annuloplasty ring sizing to pre-repair echocardiography to enable implementation of objective planning tools with improved options for risk stratification and prediction for mitral valve repair failure.

Conclusion

Within the scope of this paper, we showed the potential effect of different isolated sizing strategies on the mitral valve geometry. Using the height of the AML to find the optimal size of the annuloplasty ring led to the maximal reduction in mitral valve orifice area and hypothetically to the maximal increase in the coaptation area between AML and PML when compared to the other sizing methods. In contrast, the use of the AML surface area resulted in a smaller reduction in the mitral valve orifice area and a smaller increase in the coaptation area between the leaflets in theory. Sizing the mitral valve annuloplasty ring according to the intercommissural distance led to the smallest reduction in mitral valve orifice. To sum up, our results suggest that the concordance between the strategies is very limited and that they should be rather used in a combined way to make an adequate decision for the patient.

Abbreviations

3D TEE:

Real-time three-dimensional transesophageal echocardiography

AML:

Anterior mitral leaflet

CE ring:

Carpentier-Edwards Physio II ring

MITK:

Medical imaging interaction toolkit

PML:

Posterior mitral leaflet

SAM:

Systolic anterior movement

References

  1. Flameng W, Herijgers P, Bogaerts K. Recurrence of mitral valve regurgitation after mitral valve repair in degenerative valve disease. Circulation. 2003;107:1609–13.

    Article  PubMed  Google Scholar 

  2. Magne J, Sénéchal M, Mathieu P, Dumesnil JG, Dagenais F, Pibarot P. Restrictive annuloplasty for ischemic mitral regurgitation may induce functional mitral stenosis. J Am Coll Cardiol. 2008;51:1692–701.

    Article  PubMed  Google Scholar 

  3. Carpentier AF, Lessana A, Relland JY, et al. The “physio ring”: an advanced concept in mitral valve annuloplasty. Ann Thorac Surg. 1995;60:1177–86.

    Article  CAS  PubMed  Google Scholar 

  4. Bothe W, Miller DC, Doenst T. Sizing for mitral annuloplasty: where does science stop and voodoo begin? Ann Thorac Surg. 2013;95:1475–83.

    Article  PubMed  Google Scholar 

  5. Adams DH, Anyanwu AC, Rahmanian PB, et al. Large annuloplasty rings facilitate mitral valve repair in Barlow’s disease. Ann Thorac Surg. 2006;82:2096–101.

    Article  PubMed  Google Scholar 

  6. Carpentier A. Cardiac valve surgery—the“french correction”. J Thorac Cardiovasc Surg. 1983;86:323–37.

    CAS  PubMed  Google Scholar 

  7. Chen FY, Cohn LH. Mitral valve repair. In: Cohn LH, Edmunds LH, editors. Cardiac surgery in the adult. 2nd ed. New York: McGraw-Hill; 2008. p. 1013–29.

    Google Scholar 

  8. Chung CH, Kim JB, Choo SJ, et al. Long-term outcomes after mitral ring annuloplasty for degenerative mitral regurgitation: Duran ring versus Carpentier-Edwards ring. J Heart Valve Dis. 2007;16:536–45.

    PubMed  Google Scholar 

  9. Cook RC, Nifong LW, Lashley GG, et al. Echocardiographic measurements alone do not provide accurate non-invasive selection of annuloplasty band size for robotic mitral valve repair. J Heart Valve Dis. 2006;15:524–7.

    PubMed  Google Scholar 

  10. Perier P, Hohenberger W, Lakew F, et al. Toward a new paradigm for the reconstruction of posterior leaflet prolapse: midterm results of the “respect rather than resect” approach. Ann Thorac Surg. 2008;86:718–25.

    Article  PubMed  Google Scholar 

  11. Borghetti V, Campana M, Scotti C, et al. Biological versus prosthetic ring in mitral-valve repair: enhancement of mitral annulus dynamics and left-ventricular function with pericardial annuloplasty at long term. Eur J Cardiothorac Surg. 2000;17:431–9.

    Article  CAS  PubMed  Google Scholar 

  12. Chang BC, Youn YN, Ha JW, et al. Long-term clinical results of mitral valvuloplasty using flexible and rigid rings: a prospective and randomized study. J Thorac Cardiovasc Surg. 2007;133:995–1003.

    Article  PubMed  Google Scholar 

  13. Dreyfus GD, Souza Neto O, Aubert S. Papillary muscle repositioning for repair of anterior leaflet prolapse caused by chordal elongation. J Thorac Cardiovasc Surg. 2006;132:578–84.

    Article  PubMed  Google Scholar 

  14. Fattouch K, Guccione F, Sampognaro R, et al. POINT. Efficacy of adding mitral valve restrictive annuloplasty to coronary artery bypass grafting in patients with moderate ischemic mitral valve regurgitation: a randomized trial. J Thorac Cardiovasc Surg. 2009;138:278–85.

    Article  PubMed  Google Scholar 

  15. Rankin JS, Orozco RE, Addai TR, et al. Several new considerations in mitral valve repair. J Heart Valve Dis. 2004;13:399–409.

    PubMed  Google Scholar 

  16. Braun J, Bax JJ, Versteegh MI, et al. Preoperative left ventricular dimensions predict reverse remodeling following restrictive mitral annuloplasty in ischemic mitral regurgitation. Eur J Cardiothorac Surg. 2005;27:847–53.

    Article  PubMed  Google Scholar 

  17. Braun J, van de Veire NR, Klautz RJ, et al. Restrictive mitral annuloplasty cures ischemic mitral regurgitation and heart failure. Ann Thorac Surg. 2008;85:430–7.

    Article  PubMed  Google Scholar 

  18. Geidel S, Lass M, Schneider C, et al. Downsizing of the mitral valve and coronary revascularization in severe ischemic mitral regurgitation results in reverse left ventricular and left atrial remodeling. Eur J Cardiothorac Surg. 2005;27:1011–6.

    Article  PubMed  Google Scholar 

  19. Geidel S, Lass M, Krause K, et al. Early and late results of restrictive mitral valve annuloplasty in 121 patients with cardiomyopathy and chronic mitral regurgitation. Thorac Cardiovasc Surg. 2008;56:262–8.

    Article  CAS  PubMed  Google Scholar 

  20. Geidel S, Lass M, Ostermeyer J. Restrictive mitral valve annuloplasty for chronic ischemic mitral regurgitation: a 5-year clinical experience with the physio ring. Heart Surg Forum. 2008;11:E225–30.

    Article  PubMed  Google Scholar 

  21. Gelsomino S, Lorusso R, Capecchi I, et al. Left ventricular reverse remodeling after undersized mitral ring annuloplasty in patients with ischemic regurgitation. Ann Thorac Surg. 2008;85:1319–30.

    Article  PubMed  Google Scholar 

  22. Nolden M, Zelzer S, Seitel A, Wald D, Müller M, Franz AM, Maleike D, Fangerau M, Baumhauer M, Maier-Hein L, Maier-Hein KH, Meinzer HP, Wolf I. The medical imaging InteractionToolkit: challenges and advances : 10 years of open-source development. Int J Comput Assist Radiol Surg. 2013;8:607–20.

    Article  PubMed  Google Scholar 

  23. Wolf I, Vetter M, Wegner I, Bottger T, Nolden M, Schobinger M, Hastenteufel M, Kunert T, Meinzer HP. The medical imaging interaction toolkit. Med Image Anal. 2005;9(6):594–604.

    Article  PubMed  Google Scholar 

  24. Graser B, Wald D, Al-Maisary S, Grossgasteiger M, De Simone R, Meinzer H-P, Wolf I. Using a shape prior for robust modeling of the mitral annulus on 4D ultrasound data. Int J Comput Assist Radiol Surg. 2014;9(4):635–44.

    CAS  PubMed  Google Scholar 

  25. Ender J, Eibel S, Mukherjee C, Mathioudakis D, Borger MA, Jacobs S, Mohr FW, Falk V. Prediction of the annuloplasty ring size in patients undergoing mitral valve repair using real-time three-dimensional transoesophageal echocardiography. Eur J Echocardiogr. 2011;12(6):445–53.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  26. Schoch N, Kißler F, Stoll M, Engelhardt S, De Simone R, Wolf I, Bendl R, Heuveline V. Comprehensive patient-specific information preprocessing for cardiac surgery simulations. Int J Comput Assist Radiol Surg. 2016;11:1051–9.

    Article  CAS  PubMed  Google Scholar 

  27. Al-Maisary S, Graser B, Engelhardt S, Wolf I, Karck M, De Simone R. The geometrical effect of different annuloplasty rings on mitral valve annulus. J Cardiovasc Surg (Torino). 2016 [Epub ahead of print].

  28. Bothe W, Kvitting JPE, Swanson JC, Hartnett S, Ingels NB, Miller DC. Effects of different annuloplasty rings on anterior mitral leaflet dimensions. J Thorac Cardiovasc Surg. 2010;139:1114–22.

    Article  PubMed  PubMed Central  Google Scholar 

  29. Engelhardt S, Wolf I, Al-Maisary S, Schmidt H, Meinzer H-P, Karck M, De Simone R. Intraoperative quantitative mitral valve analysis using optical tracking technology. Annals of Thoracic Surgery. 2016;101:1950–6.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was carried out with support of the German Research Foundation (DFG) as part of project B01, SFB/TRR 125, Cognition-Guided Surgery.

Availability of data and materials

Data are available as a Additional file 1.

Authors’ contributions

SAl-M designed the study and drafted the article, SE performed data analysis and drafted the article, BG performed data analysis, IW performed critical revisions of the article, MK performed critical revisions of the article. RDS performed critical revisions of the article. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Not applicable.

Ethics approval and consent to participate

The study protocol was approved by the institutional research ethics board.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Sameer Al-Maisary.

Additional file

Additional file 1:

Measurements of mitral annulus dimensions extracted from Mitralyzer software. (XLSX 42 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Al-Maisary, S., Engelhardt, S., Graser, B. et al. Computer-based comparison of different methods for selecting mitral annuloplasty ring size. J Cardiothorac Surg 12, 8 (2017). https://doi.org/10.1186/s13019-017-0571-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-017-0571-y

Keywords