Heartmate 3 fully magnetically levitated left ventricular assist device for the treatment of advanced heart failure –1 year results from the Ce mark trial

Abstract

Background

The HeartMate 3 Left Ventricular Assist System (LVAS) (St. Jude Medical Inc., St Paul, MN) with full magnetic levitation allows for wide and consistent blood flow paths and an artificial pulse designed for enhanced hemocompatibility. The HeartMate 3 received market approval in the European Union in 2015 following completion of a multicenter study. After reaching the 6-month study endpoint, patients continue to be followed for 2 years with the 1-year results presented herein.

Methods

A prospective, non-randomized study included adults with advanced heart failure and ejection fraction (EF) ≤ 25%, cardiac index (CI) ≤ 2.2 L/min/m2 while not on inotropes, or inotrope dependent, or on optimal medical management for 45/60 days.

Results

Fifty patients—54% bridge to transplant (BTT) and 46% destination therapy (DT)—were enrolled and implanted with the HeartMate 3. At baseline, 92% of the patients were INTERMACS profiles 2–4, with cardiac index 1.8 + 0.5 L/min/m2 and 58% were supported with inotropes. At 1 year, 74% of the patients remain on support, 18% expired, 6% transplanted, and 2% explanted. The adverse events include 12% gastrointestinal bleeding, 16% driveline infections, 18% strokes, and 2% outflow graft thrombosis. There was no hemolysis, pump thrombosis or pump malfunction through 1 year. The six-minute walk test distance increased from a mean of 273 m to 371 m (P <0.0001). EQ-5D quality-of-life score increased from a mean of 52.7 to 70.8 (P = 0.0006).

Conclusions

The 1-year HeartMate 3 LVAS results show survival and adverse-event profile are similar to other approved devices, with no pump thrombosis or pump failure. Patient’s functional status and quality of life significantly improved over time.

Trial registration

Clinicaltrials.gov registration number: NCT02170363. Registered June 19, 2014.

Peer Review reports

Background

Circulatory support with a continuous-flow left ventricular assist device (CF-LVAD) restores circulation, preserves organ function, improves functional status and quality of life, and extends survival for patients with refractory advanced stage heart failure [16]. Limitations of current CF-LVAD technology is primarily the result of untoward hematologic responses to the device’s artificial materials and the high shear forces imparted on blood components by the pump [7]. Hemolysis, platelet activation, and cleaving of von Willebrand factor multimeres often occur during CF-LVAD support and are important contributing factors to thrombosis, stroke, and gastrointestinal (GI) bleeding—adverse events that are the primary limitations of CF-LVAD support [8]. Advancing outcomes of CF-LVAD-supported patients entails novel designs that optimize the blood-biomaterial interface and produce blood flow with low shear force for nominal effects on blood components.

The HeartMate 3 Left Ventricular Assist System (LVAS) (St. Jude Medical Inc., St. Paul, MN) was designed for enhanced hemocompatibility by the integration of a fully magnetically levitated rotor for frictionless movement, wide blood flow gaps through the pump for lower shear stress, and textured blood-contacting inner and outer surfaces for establishing a tissue interface with blood [9, 10]. An artificial pulse allows for more complete washing of pump surfaces eliminating stasis as a cause of pump thrombosis and possibly other clinical benefits related to the pulse pressure. Clinical studies evaluating the HeartMate 3 were initiated in June of 2014 at 10 centers in Europe, Australia, Canada, and Kazakhstan [11]. Fifty patients who met the criteria for refractory advanced stage heart failure received the HeartMate 3 in a trial designed to include both bridge to transplant and destination therapy patients [10]. Results of the study demonstrated the performance goal at the 6-month study endpoint was exceeded, with 92% of patients alive with no instances of hemolysis, pump thrombosis or pump failure. Market approval in the European Union was obtained on October 8, 2015. The purpose of this report is to present the 1 year results of the study, and planned follow-up of 2-years for this study group is ongoing [10].

Methods

The clinical trial protocol, including patient criteria for enrollment, study endpoints, HeartMate 3 LVAS description, and statistical methods have been previously published [10]. In brief, this single-arm, prospective, nonrandomized clinical trial had a primary endpoint of survival at 6 months compared with a performance goal of 88% derived using data from matched subjects in the INTERMACS (Interagency Registry for Mechanically Assisted Circulatory Support) registry. Patients with ongoing HeartMate 3 support after 6-months will be followed for 24 months. Patients were adults with New York Heart Association (NYHA) class IIIB or IV, or American College of Cardiology/American Heart Association Stage D heart failure, EF < 25%, CI < 2.2 L/min/m2 without inotropes, or were inotrope-dependent on optimal medical management, or listed for heart transplant. All patients meeting the study criteria were included regardless of the indication for LVAS support: BTT and DT. Follow-up assessments include quality of life (QOL; European Quality of Life Questionnaire 5 level); functional status (six-minute walk test, NYHA functional class); adverse event rates; incidence of device malfunction, reoperation, rehospitalization, and survival free of debilitating stroke (modified Rankin Score > 3). The INTERMACS adverse event definitions were used during the trial [12].

Results

Patients

The 50 patients were enrolled in this clinical trial from June 2014 to November 2014, with the 6-month follow-up completed on May 26, 2015, and the 1-year follow-up completed on November 26, 2015. Details of the study design and results from the first 6 months have been previously reported (10). Table 1 lists the key demographic parameters at the time of implantation for the study group. The indication for LVAS support was nearly equally distributed between BTT (54%) and DT (46%). There were no patients classified as INTERMACS profile 1 or 7, while the majority of patients were either profile 3 (42%) or 4 (40%).

Table 1 Baseline demographics for the 50 trial patients

Outcomes at 1 year

At the 1-year follow-up, 37 (74%) patients had ongoing support. Three patients (6%) underwent transplant following 50, 101, and 231 days of support, with all alive and well at follow-up. The Kaplan-Meier survival at the 12-month time point was 81 ± 6% (Fig. 1). One patient (2%) had the device explanted on support day 185 due to sepsis and an abdominal wall abscess. There were a total of 9 deaths (18%): 4 in the first 6-months and 5 in the second 6-month follow-up time. The causes of deaths in the first 6-months were anaphylaxis, cardiac arrest, suicide and renal failure [10]. Deaths occurring during months 6 to 12 were caused by Merkel Cell carcinoma (day 203), ischemic stroke (day 212), intracerebral bleeding with international normalized ratio (INR) > 6 (day 232), and 2 cases of chronic infection with multiple organ failure (days 277 and 319).

Fig. 1
figure 1

Kaplan-Meier survival to 1-year after implantation

Forty-three (86%) patients were discharged from the hospital with LVAS support. Seven patients were not discharged due to death during the index hospitalization (5) and heart transplantation (2). Twenty-eight patients (56%) had a total 66 readmissions to the hospital due to various adverse events (51; 77%), routine or scheduled testing (5), low flow alarm (2), driveline infection (1), ICD therapy (1), epistaxis (1), automobile accident (1), broken arm from fall (1), transplant evaluation (1), anticoagulation adjustment (1), and suspected device malfunction (1).

Adverse events

Adverse events are presented in Table 2 by the interval of occurrence (days 0 to 180 or days 181 to 365) and the total for the 1-year follow-up. Bleeding and infection were the predominate events overall, occurring in 44% and 36% of patients, respectively. Bleeding events requiring reoperation or blood transfusion were observed mostly early after implantation. The overall bleeding event rate declined from 38% in the first 6 months to 18% in the second 6-month period. The total number of patients with GI bleeding was 6 (12%), with only 2 patients (5%) experiencing this event during the second 6 months of support. The freedom from GI bleeding event was 89.6% at 6 months and 87.1% at 1 year (Fig. 2). The number of total infection events were fewer during the second 6 months (28 vs 10), with 3 (7%) having driveline infection and 2 (5%) with sepsis.

Table 2 Adverse events
Fig. 2
figure 2

Freedom from gastrointestinal (GI) bleed

Stroke events occurred in 6 (12%) patients in the first 6-months and in 3 (6%) patients in the second 6 months. The freedom from stroke was 87.7% at 6 months and 81.2% at 1-year (Fig. 3). Of those patients with stroke in the second 6 months, one patient experienced ischemic stroke with a modified Rankin Score (MRS) > 3 on support day 209 following rapid onset of sepsis; computed tomography scan identified outflow thrombosis, which embolized during transport to the operating room. Two patients had hemorrhagic strokes, with MRS > 3: one on support day 231 that was associated with an INR > 6, and another on day 309 that was associated with a chronic driveline infection.

Fig. 3
figure 3

Freedom from stroke

During the 1-year follow-up time, there were no incidences of hemolysis, pump thrombosis or pump malfunction. Other neurologic events and right heart failure were not observed in the second 6-month follow-up.

Functional status and quality of life

Patients who were able to complete the six-minute walk test before implantation and at 1, 3, 6, and 12 months showed a statistically significant improvement in walk distance at each interval (Fig. 4). At 1 year, the median distance walked was 383 m, which was an improvement of 219 m over the baseline values. The NYHA class was significantly improved (p <0.0001), with 89% of patients being classified as I or II at 1-year time point compared to 0% prior to implant (Fig. 5). Patient assessment of their QOL by the EuroQol visual analog score showed significant improvement at 3, 6, and 12 months (p <0.001) (Fig. 6).

Fig. 4
figure 4

Six-minute walk test data paired for patients with results at baseline and at the specified interval

Fig. 5
figure 5

Change in the percent of patients in each NYHA classification at each interval

Fig. 6
figure 6

Visual analog score data paired for patients with results at baseline and at the specified interval

Discussion

This follow-up evaluation of heart failure patients supported by the HeartMate 3 for BTT and DT indicates sustained improvement in physical functioning and QOL out to 1 year. Adverse event rates tended to be lower in the second 6 months and there were no cases of hemolysis, pump thrombosis, or pump failure in the year of support. These results demonstrate comparable reliability and safety of the HeartMate 3 at 1 year of support as compared with currently available CF-LVADs.

The occurrence of GI bleeding during LVAS support is an important concern because of the need for rehospitalization and the considerable increase in the cost of care. The reported rates of GI bleeding range from 15%[13] to as high as 40%,[14] with the majority of studies reporting rates in the range of 20–30% [1520]. Loss of von Willebrand multimers due to high shear stress and decreased arterial pulsatility are believed to be the principle characteristics of CF-LVADs that cause GI bleeding [2124]. The 1-year rate of GI bleeding in this study was 12%, with only 5% occurring during the second half of the year. This lower rate of GI bleeding appears to be due to better preservation of von Willebrand multimeres that has been observed in patients supported by the HeartMate 3 [25].

Pump thrombosis has become a considerable concern in recent years due to an increase in incidence and the considerable morbidity associated with the event [26, 27]. Most patients survive this event by undergoing device exchange, or in a few cases thrombolytic therapy is successful. Pump exchange can be performed safely; however, there is some mortality due to major surgery and the hospital time increases cost considerably [2831]. The rate of pump thrombosis for patients implanted with the HeartMate II LVAS (St. Jude Medical Inc., St. Paul MN) is between 6% and 12% at one year and for the HeartWare HVAD (HeartWare International Inc., Framingham, MA), the reported rate is 8% [26, 32, 33]. An important observation from this relatively small study is that there were no cases of HeartMate 3 pump thrombosis, and no pump failures, and thus no patients required pump exchange.

The overall rate of stroke at 18%, with near equal distribution between ischemic (10%) and hemorrhagic (8%) stroke, was not expected given the absence of hemolysis and pump thrombosis indicating that these events may be related other factors including blood pressure and other variables. In comparison, in recent publications the stroke rate for patients implanted with the HeartMate II LVAS (St. Jude Medical Inc., St. Paul MN) is between 10.9% and 12.1% (6 month and 2 year follow-up rate, respectively) [34, 35] and for the HeartWare HVAD (HeartWare International Inc., Framingham, MA), the 2 year follow-up reported rate is 29.7% [35]. Six (12%) of the stroke events occurred within the first 6 months and 3 (7%) occurred in the second half of the follow-up time. Factors associated with the stroke events were procedural (3), infection (3), INR > 6 (1), multiple organ failure (1), and 1 unidentified. Strokes associated with procedures were 1) difficulty placing the inflow conduit, 2) anaphylaxis from contrast media for lung CT scan, 3) following transcather aortic valve placement [10]. All patients were receiving warfarin anticoagulation, with a target INR range of 2.0–3.0, and aspirin 81–100 mg daily. Procedural-related stroke events are controllable and infection reduction efforts are continuous. Along with refined anticoagulation protocols that should be revealed in ongoing studies, there should be optimism that stroke events will decline in the future. Other neurologic events encountered were minimal, with 2 (4%) transient ischemic attacks (TIA) and 2 (4%) seizures; all which occurred within the first 6 months.

The 30-day, 6-month, and 12-month survival rates for this cohort were 98, 92, and 81%, respectively. Careful consideration must be given when comparing these results to other clinical studies with durable CF-LVADs because of the mix of BTT and DT indications. The 1-year survival rate for BTT studies is 85%[35]; for DT, the rate is 74% [36]. In a recent report of a study involving all patients receiving durable support (N = 5942) from the International Society for Heart and Lung Transplantation Mechanically Assisted Circulatory Support (IMACS) registry, the 6-month and 1-year survival rates were 86 and 80%, respectively [37]. Although any comparison to other studies is imprecise, the survival rates for this study are similar to other experiences.

Limitations

The foremost limitations of this study was the non-randomized and non-controlled design and the relatively small number of patients in the study. Comparison to other clinical studies is not equivalent due to the mix of BTT and DT indications. With implantations at 10 different centers and in different parts of the world, patient care practices may vary by institutional preferences. All sites participating in the trial received the same training and support, and each site had similar longstanding experience with mechanical circulatory support systems.

Conclusions

The survival rate is acceptable in this first in human series of patients with the HeartMate 3 LVAS. Major complications such as bleeding, infection, stroke, and right heart failure declined in the second 6 months of follow-up. Improvements in NYHA class, six-minute walk distance, and QOL measures demonstrate excellent circulatory support.[38] The absence of hemolysis and pump thrombosis for the 1-year follow-up is encouraging. The results observed in this study mirror that of a large randomized clinical trial just recently completed in the United States [34, 39]. In this trial that compared the HeartMate II and the HeartMate 3 in an “all-comers” design, reoperation due to pump thrombosis favored the new centrifugal flow HeartMate 3. Both HeartMate 3 studies indicate that there appears to be enhanced durability and hemocompatibility for long-term support with the HeartMate 3.

Abbreviations

BTT:

Bridge to transplant

CF-LVAD:

Continuous-flow left ventricular assist device

CI:

Cardiac index

DT:

Destination therapy

EF:

Ejection fraction

GI:

Gastrointestinal

IMACS:

International society for heart and lung transplantation mechanically assisted circulatory support

INR:

International normalized ratio

INTERMACS:

Interagency registry for mechanically assisted circulatory support

LVAS:

Left ventricular assist system

NYHA:

New York Heart Association

QOL:

Quality of life

RVAD:

right ventricular assist device

References

  1. Pagani FD, Miller LW, Russell SD, et al. Extended mechanical circulatory support with a continuous-flow rotary left ventricular assist device. J Am Coll Cardiol. 2009;54:312–21.

    Article  PubMed  Google Scholar 

  2. Slaughter MS, Rogers JG, Milano CA, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361:2241–51.

    Article  CAS  PubMed  Google Scholar 

  3. John R, Naka Y, Smedira NG, et al. Continuous flow left ventricular assist device outcomes in commercial use compared with the prior clinical trial. Ann Thorac Surg. 2011;92:1406–13. discussion 13.

    Article  PubMed  Google Scholar 

  4. Starling RC, Naka Y, Boyle AJ, et al. Results of the post-U.S. Food and drug administration-approval study with a continuous flow left ventricular assist device as a bridge to heart transplantation a prospective study using the INTERMACS (interagency registry for mechanically assisted circulatory support). J Am Coll Cardiol. 2011;57:1890–8.

    Article  PubMed  Google Scholar 

  5. Slaughter MS, Pagani FD, McGee EC, et al. HeartWare ventricular assist system for bridge to transplant: combined results of the bridge to transplant and continued access protocol trial. J Heart Lung Transplant. 2013;32:675–83.

    Article  PubMed  Google Scholar 

  6. Strueber M, Larbalestier R, Jansz P, et al. Results of the post-market registry to evaluate the HeartWare left ventricular assist system (ReVOLVE). J Heart Lung Transplant. 2014;33:486–91.

    Article  PubMed  Google Scholar 

  7. Uriel N, Adatya S, Mehra MR. Evolution in mechanical circulatory support: Are We at the precipice of a disruptive innovation? J Am Coll Cardiol. 2015;66:2590–3.

    Article  PubMed  Google Scholar 

  8. Pagani FD. Adverse event burden and mechanical circulatory support: looking toward the future. J Thorac Cardiovasc Surg. 2016;151:10–2.

    Article  PubMed  Google Scholar 

  9. Bourque K, Cotter C, Dague C, Harjes D, Dur O, Duhamel J, Spink K, Walsh K, Burke E. Design rationale and preclinical evaluation of the HeartMate 3 left ventricular assist system for hemocompatibility. ASAIO J. 2016;62:375–83.

    Article  CAS  PubMed  Google Scholar 

  10. Netuka I, Sood P, Pya Y, et al. Fully magnetically levitated left ventricular assist system for treating advanced HF: a multicenter study. J Am Coll Cardiol. 2015;66:2579–89.

    Article  PubMed  Google Scholar 

  11. Schmitto JD, Hanke JS, Rojas SV, Avsar M, Haverich A. First implantation in man of a new magnetically levitated left ventricular assist device (HeartMate III). J Heart Lung Transplant. 2015;34:858–60.

    Article  PubMed  Google Scholar 

  12. Manual of operations; Protocol: Appedix A Adverse event definitions. https://www.uab.edu/medicine/intermacs/appendices/app-a-5-0.

  13. Goldstein DJ, Aaronson KD, Tatooles AJ, et al. Gastrointestinal bleeding in recipients of the HeartWare ventricular assist system. JACC Heart Fail. 2015;3:303–13.

    Article  PubMed  Google Scholar 

  14. Stern DR, Kazam J, Edwards P, et al. Increased incidence of gastrointestinal bleeding following implantation of the HeartMate II LVAD. J Card Surg. 2010;25:352–6.

    Article  PubMed  Google Scholar 

  15. Morgan JA, Paone G, Nemeh HW, et al. Gastrointestinal bleeding with the HeartMate II left ventricular assist device. J Heart Lung Transplant. 2012;31:715–8.

    Article  PubMed  Google Scholar 

  16. Stulak JM, Lee D, Haft JW, et al. Gastrointestinal bleeding and subsequent risk of thromboembolic events during support with a left ventricular assist device. J Heart Lung Transplant. 2014;33:60–4.

    Article  PubMed  Google Scholar 

  17. Holley CT, Harvey L, Roy SS, et al. Gastrointestinal bleeding during continuous-flow left ventricular assist device support is associated with lower rates of cardiac transplantation. ASAIO J. 2015;61:635–9.

    Article  PubMed  Google Scholar 

  18. Lopilato AC, Doligalski CT, Caldeira C. Incidence and risk factor analysis for gastrointestinal bleeding and pump thrombosis in left ventricular assist device recipients. Artif Organs. 2015;39:939–44.

    Article  CAS  PubMed  Google Scholar 

  19. Tsiouris A, Paone G, Nemeh HW, et al. Lessons learned from 150 continuous-flow left ventricular assist devices: a single institutional 7 year experience. ASAIO J. 2015;61:266–73.

    Article  PubMed  Google Scholar 

  20. Dell’Aquila AM, Schneider SR, Schlarb D, et al. Initial clinical experience with the HeartWare left ventricular assist system: a single-center report. Ann Thorac Surg. 2013;95:170–7.

    Article  PubMed  Google Scholar 

  21. Meyer AL, Malehsa D, Budde U, Bara C, Haverich A, Strueber M. Acquired von willebrand syndrome in patients with a centrifugal or axial continuous flow left ventricular assist device. JACC Heart Fail. 2014;2:141–5.

    Article  PubMed  Google Scholar 

  22. Uriel N, Pak SW, Jorde UP, et al. Acquired von willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation. J Am Coll Cardiol. 2010;56:1207–13.

    Article  PubMed  Google Scholar 

  23. Moazami N, Dembitsky WP, Adamson R, et al. Does pulsatility matter in the era of continuous-flow blood pumps? J Heart Lung Transplant. 2015;34:999–1004.

    Article  PubMed  Google Scholar 

  24. Wever-Pinzon O, Selzman CH, Drakos SG, et al. Pulsatility and the risk of nonsurgical bleeding in patients supported with the continuous-flow left ventricular assist device HeartMate II. Circ Heart Fail. 2013;6:517–26.

    Article  CAS  PubMed  Google Scholar 

  25. Netuka I, Kvasnicka T, Kvasnicka J, Hrachovinova I, Ivak P, Marecek F, Bilkova J, Malikova I, Jancova M, Maly J, Sood P, Sundareswaran KS, Connors JM, Mehra MR. Evaluation of von willebrand factor with a fully magnetically levitated centrifugal continuous-flow left ventricular assist device in advanced heart failure. J Heart Lung Transplant. 2016;35:860–7.

    Article  PubMed  Google Scholar 

  26. Starling RC, Moazami N, Silvestry SC, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370:33–40.

    Article  CAS  PubMed  Google Scholar 

  27. Mehra MR, Stewart GC, Uber PA. The vexing problem of thrombosis in long-term mechanical circulatory support. J Heart Lung Transplant. 2014;33:1–11.

    Article  PubMed  Google Scholar 

  28. Levin AP, Uriel N, Takayama H, et al. Device exchange in HeartMate II recipients: long-term outcomes and risk of thrombosis recurrence. ASAIO J. 2015;61:144–9.

    Article  PubMed  Google Scholar 

  29. McCarthy FH, Kobrin D, Rame JE, et al. Increasing frequency of left ventricular assist device exchanges in the united states. Ann Thorac Surg. 2015;100:1660–4. discussion 1665.

    Article  PubMed  Google Scholar 

  30. Soleimani B, Stephenson ER, Price LC, El-Banayosy A, Pae WE. Clinical experience with sternotomy versus subcostal approach for exchange of HeartMate II left ventricular assist device. Ann Thorac Surg. 2015;100:1577–80.

    Article  PubMed  Google Scholar 

  31. Oezpeker C, Zittermann A, Ensminger S, et al. Systemic thrombolysis versus device exchange for pump thrombosis management: a single center experience. ASAIO J. 2016;62:246–51.

    Article  PubMed  Google Scholar 

  32. Najjar SS, Slaughter MS, Pagani FD, et al. An analysis of pump thrombus events in patients in the HeartWare ADVANCE bridge to transplant and continued access protocol trial. J Heart Lung Transplant. 2014;33:23–34.

    Article  PubMed  Google Scholar 

  33. Kirklin JK, Naftel DC, Pagani FD, et al. Pump thrombosis in the thoratec HeartMate II device: an update analysis of the INTERMACS registry. J Heart Lung Transplant. 2015;34:1515–26.

    Article  PubMed  Google Scholar 

  34. Mehra MR, Naka Y, Uriel N, Goldstein DJ, Cleveland Jr JC, Colombo PC, Walsh MN, Milano CA, Patel CB, Jorde UP, Pagani FD, Aaronson KD, Dean DA, McCants K, Itoh A, Ewald GA, Horstmanshof D, Long JW, Salerno C, Investigators M. A fully magnetically levitated circulatory pump for advanced heart failure. N Engl J Med. 2017;376:440–50.

    Article  PubMed  Google Scholar 

  35. Rogers JG, Pagani FD, Tatooles AJ, Bhat G, Slaughter M, Birks EJ, Boyce SW, Najjar SS, Jeevanandam V, Anderson AS, Gregoric ID, Mallidi H, Leadley K, Aaronson KD, Frazier OH, Milano CA. Intrapericardial left ventricular assist device for advanced heart failure. N Engl J Med. 2017;376:451–60.

    Article  PubMed  Google Scholar 

  36. Jorde UP, Kushwaha SS, Tatooles AJ, et al. Results of the destination therapy post-food and drug administration approval study with a continuous flow left ventricular assist device: a prospective study using the INTERMACS registry (interagency registry for mechanically assisted circulatory support). J Am Coll Cardiol. 2014;63:1751–7.

    Article  PubMed  Google Scholar 

  37. Kirklin JK, Cantor R, Mohacsi P, et al. First annual IMACS report: a global international society for heart and lung transplantation registry for mechanical circulatory support. J Heart Lung Transplant. 2016;35:407–12.

    Article  PubMed  Google Scholar 

  38. Uriel N, Adatya S, Maly J, Kruse E, Rodgers D, Heatley G, Herman A, Sood P, Berliner D, Bauersachs J, Haverich A, Želízko M, Schmitto JD, Netuka I. Clinical hemodynamic evaluation of patients implanted with a fully magnetically levitated left ventricular assist device (HeartMate 3). J Heart Lung Transplant. 2017;36:28–35.

  39. Heatley G, Sood P, Goldstein D, et al. Clinical trial design and rationale of the multicenter study of MagLev technology in patients undergoing mechanical circulatory support therapy with HeartMate 3 (MOMENTUM 3) investigational device exemption clinical study protocol. J Heart Lung Transplant. 2016;35:528–36.

    Article  PubMed  Google Scholar 

Download references

Acknowledgement

Timothy J. Myers of St. Jude Medical Inc. provided technical writing service for this manuscript.

Funding

St. Jude Medical Inc. was the sponsor of this study and participated with the investigators in the design of the study and collection, analysis, and interpretation of data and in writing the manuscript.

Availability of data and materials

All data generated or analysed during this study are included in this article.

Authors’ contributions

TK drafted and reviewed the manuscript, participated in the design of the study, reviewed the analysis and performed data collection for the study; IN reviewed the manuscript, reviewed the analysis and performed data collection for the study; JS reviewed the manuscript, reviewed the data analysis and performed data collection for the study; DZ reviewed the manuscript and performed data collection for the study; JG reviewed the manuscript and performed data collection for the study; MM reviewed the manuscript, participated in the design of the study and performed data collection for the study; FB reviewed the manuscript and performed data collection for the study; SM reviewed the manuscript and performed data collection for the study; LD drafted and reviewed the manuscript, coordinated the study conduct and design of the study; YP reviewed the manuscript and performed data collection for the study. All authors read and approved the final manuscript.

Competing interests

The authors declare that they have no competing interests.

Consent for publication

Individuals are not identified in this report; therefore, specific consent for publication was not obtained. All patients are informed that the data obtained from their participation in this research would be reported in the medical literature.

Ethics approval and consent to participate

EC approval was obtained from EC of Hannover Medical School, Germany (lead center) on April 24, 2014, Nr. 6604 MPG-LKP. Additionally, the ethical committees of each participating institution approved the research protocol. All study participants gave informed consent.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Thomas Krabatsch.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Krabatsch, T., Netuka, I., Schmitto, J.D. et al. Heartmate 3 fully magnetically levitated left ventricular assist device for the treatment of advanced heart failure –1 year results from the Ce mark trial. J Cardiothorac Surg 12, 23 (2017). https://doi.org/10.1186/s13019-017-0587-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-017-0587-3

Keywords