Skip to main content

If we should remove internal fixation devices for rib fractures?

Abstract

Background

Internal fixation for rib fractures has been widely carried out worldwide, and its surgical efficacy has been recognized. However, there is still controversy about whether implant materials need to be removed. At present, the research on this topic is still lacking at home and abroad. Therefore, in this study, the patients undergoing removal of internal fixation for rib fractures in our department within one year were followed up, to statistically analyze implant-related complications, postoperative complications and postoperative remission rate.

Methods

A retrospective analysis was conducted on 143 patients undergoing removal of internal fixation for rib fractures from 2020 to 2021 in our center. The implant-related complications, postoperative complications and postoperative remission rate of patients with internal fixation were analyzed.

Results

In this study, a total of 143 patients underwent removal of internal fixation, among which 73 suffered from preoperative implant-related complications (foreign-body sensation, pain, wound numbness, sense of tightness, screw slippage, chest tightness, implant rejection), and 70 had no post operative discomfort but asked for removal of internal fixation. The average interval between rib fixation and removal was 17 ± 9.00 (months), and the average number of removed materials was 5.29 ± 2.42. Postoperative complications included wound infection (n = 1) and pulmonary embolism (n = 1). of the 73 patients with preoperative implant-related complications, the mean postoperative remission rate was 82%. Among the 70 patients without preoperative discomfort, the proportion of discomforts after removal was 10%. No perioperative death occurred.

Conclusion

For patients with internal fixation for rib fractures, removal of internal fixation can be considered in the case of implant-related complications after surgery. The corresponding symptoms can be relieved after removal. The removal presents low complication rate, and high safety and reliability. For patients without obvious symptoms, it is safe to retain the internal fixation in the body. For the asymptomatic patients who ask for removal of internal fixation, the possible risk of complications should be fully informed before removal.

Peer Review reports

Background

With the surgical concept of internal fixation for rib fractures becoming increasingly popular, internal fixation for multiple rib fractures has been carried out more and more widely [1, 2]. Along with the increase in surgical numbers, the following problem is whether the internal fixation devices of the ribs need to be removed after a certain period of time. In the technical documents provided by manufacturers, it is suggested that the internal fixation devices should be removed at an appropriate time, and the timing and indications should be determined by clinicians, but no corresponding documentation has been provided. In the existing literature, there is almost no discussion about the removal of internal fixation devices for rib fractures. Reviewing the orthopedic literature on the removal of internal fixation devices, there are considerable differences of opinion on whether internal fixation devices need to be removed in asymptomatic patients, and high-level evidence-based medical evidence is lacking to determine whether it is necessary to remove internal fixation devices for fractures [3, 4].

In actual clinical practice, a considerable number of orthopedic internal fixation devices have been removed, but the surgical indications are still quite vague. In a Finnish study, nearly all (81%) implants inserted for fracture fixation were eventually removed [5]. In total, implant removal contributed to almost 30% of all planned orthopaedic operations [6]. The reason for this perception and behaviour mismatch remains unclear particularly considering the associated high complication rate of 20% with up to 14% infection rate, 2% nerve injury and re-fracture in 0.5% [7].

We statistically analyzed the patients undergoing removal of internal fixation devices for rib fractures in the Department of Thoracic Surgery of Shanghai Sixth People’s Hospital in the past year, in the expectation for summarizing and discussing the reasons, risks and benefits of removal of internal fixation devices for rib fractures.

Methods.

The data of patients undergoing removal of internal fixation for rib fractures from 2020 to 2021 in our department were collected for a retrospective analysis. All patients underwent internal fixation with Ni-TI claw plates and Synthes rib fixation plates.The data were sorted out with the trauma database of our department. The basic clinical characteristics of the patients were analyzed. The collected data included the reasons for removal of internal fixation, preoperative symptoms, postoperative symptom remission rate and postoperative complications. The patients were followed up for symptom improvement for 3 months. All the data were analyzed.

Results

During 2020–2021, 143 patients undergoing removal of internal fixation devices for rib fractures in our department. The basic clinical information is seen in Table 1. There were 75 males and 68 females, with an average age of 56 ± 2.32 years. Between internal fixation and removal was 17 ± 9.00 months. In the same period, a total of 554 patients underwent internal fixation for rib fractures, and the proportion of patients undergoing removal of internal fixation was about 25.81% of all patients undergoing surgery in the same period. Among them, 70 (48.95%) patients had no symptoms but asked for removal of internal fixation, and 73 (51.05%) patients suffered from implant-related complications and required removal of internal fixation.

Table 1 Basic data

A statistics was also performed on the reasons for surgical removal of internal fixation, as shown in Table 2. Among all the patients with symptoms undergoing removal, the most common symptoms were pain and foreign-body sensation in the chest (n = 59), accounting for 80.82%. Other discomforts included sense of numbness and tightness (n = 11), accounting for 15.06%. Additionally, there was 1 patient requiring removal of internal fixation due to poor wound healing caused by screw slippage, chest tightness and implants, respectively. The common complications during and after removal included re-fracture and wound infection. There were 2 cases of wound infection and 3 cases of secondary fracture caused by intraoperative bone injury. In all the patients, one developed pulmonary embolism after removal, which was improved after active treatment. No death occurred among all the included patients.

Table 2 Postoperative status of patients with preoperative symptoms

Moreover, a 3-month follow-up was carried out on the improvement in patients’ symptoms after removal. In all the symptomatic patients undergoing removal, the overall improvement was 82.19%(Table 2). Among the 70 asymptomatic patients, 2 had chest wall pain after removal, and 3 presented a sense of tightness (Table 3).

Table 3 Postoperative status of patients without preoperative symptoms

Discussion

Surgical internal fixation has been generally accepted to be the most effective treatment for multiple rib fractures or flail chest. Our previous study has also confirmed that surgical internal fixation can rapidly relieve the respiratory impairment and pain caused by rib fractures [8]. However, doctors and patients in different countries have distinct perceptions about whether the internal fixation devices need to be removed. As the surgical internal fixation for rib fractures has not been carried out for a long time, we lack reference data to support any theory.

According to our study, it was found that the patients undergoing removal of internal fixation could be divided into two groups. One group without any discomfort concerning about the implants retained in the body due to social and cultural factors, and still hoped to remove the implants. In our study, most of these patients recovered well after removal of internal fixation without obvious postoperative complications, but 2 presented chest wall pain and 3 had a sense of tightness around the wound after removal. The possible reason lies in that the bone may be damaged to a certain extent during the operation, resulting in postoperative chest pain [9]. The appearance of the sense of tightness around the wound is firstly considered to be resulted from local contracture caused by scar healing. The probability of secondary trauma will be increased in removal of internal fixation to a certain extent [10].

For the other group with discomforts, internal fixation for rib fractures would still bring some adverse effects to partial patients, including pain, foreign-body sensation and discomforts. Although the proportions were not high, they would still affect the quality of life of the patients. Based on our data, the majority of the symptomatic patients had obvious remission of their discomforts after removal of the internal fixation. Among them, some symptoms of the patients with a preoperative sense of numbness and tightness were relieved after implant removal. Therefore, we consider that this sense of numbness and tightness is not only caused by the nerve terminal injury and scar healing caused by surgery, but also related to the implantation of internal fixation materials [11]. Other similar chest tightness, foreign-body sensation and poor wound healing were significantly improved after removal of the internal fixation. Consequently, for these patients, we believe that the removal of internal fixation is still necessary and valuable.

According to our data, the incidence of removal-related complications was about 4.2%. The common complications was secondary fracture caused by wound infection and bone injury. In previous studies, it was proposed that most removal of internal fixation were performed by junior physicians, so more removal-related complications might occur. In general, we believe that the removal of internal is safe with low incidence of complications, which can be treated.

Conclusion

In conclusion, for patients with internal fixation for rib fractures, removal of internal fixation can be considered in the case of chest discomforts after surgery, including pain, sense of tightness and foreign-body sensation. The corresponding symptoms can be relieved after removal. The removal presents low complication rate, and high safety and reliability. For patients without obvious symptoms, it can be considered to retain the internal fixation in the body. For the asymptomatic patients who ask for removal of internal fixation, the possible risk of complications should be fully informed before removal.

Data Availability

Data and materials are available by e-mail.

References

  1. Sawyer E, Wullschleger M, Muller N, Muller M. Surgical Rib fixation of multiple Rib Fractures and flail chest: a systematic review and Meta-analysis. J Surg Res. 2022 Aug;276:221–34.

  2. Kong LW, Huang GB, Yi YF, Du DY. Consensus expert group. The chinese consensus for surgical treatment of traumatic rib fractures 2021 (C-STTRF 2021). Chin J Traumatol. 2021 Nov;24(6):311–9.

  3. Bhandari R, Dawadi P, Thapa Magar M, Sinha R, Kayastha N, Pratap Shah R, Babu Thapa B, Rana Magar S. Orthopedic Implant removal in the Department of Orthopedics of a Tertiary Care Centre of Nepal: a descriptive cross-sectional study. JNMA J Nepal Med Assoc. 2021 Feb;28(234):116–9.

  4. Barcak EA, Beebe MJ, Weinlein JC. The role of Implant removal in Orthopedic Trauma. Orthop Clin North Am. 2018 Jan;49(1):45–53.

  5. Busam ML, Esther RJ, Obremskey WT. Hardware removal: indications and expectations. J Am Acad Orthop Surg. 2006;14:113–20.

    Article  PubMed  Google Scholar 

  6. Acklin YP, Michelitsch C, Sommer C. Elective implant removal in symptomatic patients after internal fixation of proximal humerus fractures improves clinical outcome. BMC Musculoskelet Disord 2016 Mar 10;17:119.

  7. Acklin YP, Bircher A, Morgenstern M, Richards RG, Sommer C. Benefits of hardware removal after plating. Injury. 2018 Jun;49(Suppl 1):91–S95.

  8. Li Y, Gao E, Yang Y, Gao Z, He W, Zhao Y, Wu W, Zhao T, Guo X. Comparison of minimally invasive surgery for non-flail chest rib fractures: a prospective cohort study. J Thorac Dis. 2020 Jul;12(7):3706–14.

  9. Kellam PJ, Harrast J, Weinberg M, Martin DF, Davidson NP, Saltzman CL. Complications of hardware removal. J Bone Joint Surg Am 2021 Nov 17;103(22):2089–95.

  10. Liu CH, Yeh WL, Tsai PJ, Fan KF, Cheng HW, Chen JM. A novel implant removal technique by endoscopy. J Orthop Surg Res 2018 Apr 6;13(1):74.

  11. Choi J, Kaghazchi A, Sun B, Woodward A, Forrester JD. Systematic review and Meta-analysis of Hardware failure in Surgical stabilization of Rib Fractures: who, what, when, where, and why? J Surg Res. 2021 Dec;268:190–8.

Download references

Funding

The publication fee shall be reimbursed by the Funding(ynlc201908).

Author information

Authors and Affiliations

Authors

Contributions

Yang Li. Kaile Jiang and Tiancheng Zhao wrote the main manuscript text and Xiang Guo.Kaibin Liu prepared Tables 1, 2 and 3. Yonghong Zhao did the data analysis. All authors reviewed the manuscript.

Corresponding author

Correspondence to Tiancheng Zhao.

Ethics declarations

Ethics approval

The study was approved by Shanghai Sixth People’s Hospital ethics committee (No. 2019-138-(3))

Competing interests

The authors have no conflicts of interest to declare.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, Y., Jiang, k., Zhao, T. et al. If we should remove internal fixation devices for rib fractures?. J Cardiothorac Surg 18, 214 (2023). https://doi.org/10.1186/s13019-023-02330-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-023-02330-1

Keywords