The Carpentier-Edwards Perimount Magna mitral valve bioprosthesis: intermediate-term efficacy and durability

Abstract

Background

The Carpentier-Edwards Perimount Magna mitral valve bioprosthesis (Edwards Lifesciences, Irvine, CA) is a low-profile version of the earlier Perimount valve that uses the ThermaFix process for enhanced calcium removal. The Magna valve has been in use since 2008, yet no publication, until now, has verified its intermediate-term safety and efficacy.

Methods

From 2008 through 2011 (our 4-year study period), 70 Magna valves were implanted in the mitral position at a single institution (the Cleveland Clinic). Echocardiograms were prospectively interpreted. For this study, we reviewed patients’ charts; endpoints included hemodynamic measurements, in-hospital morbidity and mortality, valve-related events, resource utilization, and 5-year survival rates.

Results

The mean patient age was 68 years; 43 % of the patients had New York Heart Association (NYHA) class III or IV disease, and 51.4 % had moderately severe, or worse, mitral regurgitation (MR). For 43 % of the patients, the Magna valve implantation was a reoperation. For 83 %, the Magna valve implantation also included a concomitant cardiac procedure. The median survival rate was 4.7 years and 90 % of patients were free from significant structural valve degeneration at 5 years. Preoperative atrial fibrillation, ischemic MR, intraaortic balloon pump placement, cardiogenic shock, cardiac arrest, and renal failure were associated with increased mortality. Right ventricular systolic pressure decreased from 50 mmHg preoperatively to 40 mmHg postoperatively, according to our matched-pair analysis (P = 0.003). Per their final echocardiogram during our study period, 98 % of surviving patients had trivial or no MR, one patient had mild MR, and one patient had severe MR.

Conclusions

Our 5-year experience indicates that the Magna valve offers excellent intermediate-term durability and substantial echocardiographic improvement; its low-profile design make it ideal for reoperations and for concomitant cardiac procedures, including valve replacement.

Peer Review reports

Background

Each year, more than 20,000 mitral valve operations are reported to the Society of Thoracic Surgeons database [1]. Despite the increased emphasis on valve repair, at least 30 % of patients with mitral valve disease still undergo valve replacement [24]. Reasons include extensive comorbidities, previous valve operations, complex jets, and mitral stenosis [3]. The ideal replacement valve would be durable, would not require anticoagulation, and would be small enough to avoid distortion of the mitral annulus while preserving left ventricular geometry.

A mitral valve bioprosthesis avoids anticoagulation, but is associated with a higher reoperation rate than a mechanical valve. For patients older than 70 years and for patients with a contraindication to anticoagulation, a mitral valve bioprosthesis is the preferred replacement option [1, 58]. Several are commercially available and approved by the U.S. Food and Drug Administration (FDA). Among the most popular are the St. Jude Medical Biocor (St. Jude Medical, St. Paul, MN) and the Carpentier-Edwards Perimount mitral valve bioprosthesis (Edwards Lifesciences, Irvine, CA). With both of those valves, the published short- and long-term hemodynamic and clinical data have suggested good outcomes [7, 912].

In 2008, Edwards Lifesciences modified that initial Perimount model to provide a lower profile and to enhance removal of calcium-binding sites through its ThermaFix process. This modified model is referred to as the Carpentier-Edwards Perimount Magna mitral valve bioprosthesis, or the Magna valve for short. It protrudes less into the left ventricular outflow tract (LVOT) than its predecessor, so it is appealing for patients with small ventricles undergoing multiple valve procedures or reoperations. It is also predicted to have less structural valve deterioration, thanks to the ThermaFix process.

The Magna valve, building on the proven durability of the initial Perimount model, has now been implanted in centers around the United States. Yet no publication, until now, has verified its short- and intermediate-term safety and efficacy. Herein, we report our 5-year outcomes with the use of the Magna valve in 70 patients implanted between 2008 through 2011 at a single center (the Cleveland Clinic)—including short-term in vivo echocardiographic data, which can be used as a reference for outcomes analyses and for future valve modifications.

Methods

Study population

For our study, we queried the Cardiovascular Information Registry (Cleveland Clinic), a prospective database approved for use in research by the Institutional Review Board. Included in our study population were patients who underwent mitral valve replacement with the Magna valve from 2008 through 2011 (our 4-year study period) (Fig. 1). Excluded were patients who underwent mitral valve repairs or replacements with other types of valves.

Fig. 1
figure 1

a. Original Carpentier-Edwards Perimount design. b. Magna valve based on the original design but now with a lower profile and Therma fix processing

In general the Cleveland Clinic prefers mitral valve repair whenever possible. When replacement is needed, we have primarily used either the St. Jude Medical Biocor or the Carpentier-Edwards Perimount mitral valve bioprosthesis; however, since 2008, our experience with the Magna valve has been gradually accumulating.

After querying the Cardiovascular Information Registry, we reviewed patients’ electronic medical records and entered their clinical data into our own working database for analysis. This database was approved for use by the Cleveland Clinic’s Institutional Review Board, which waived the need for consent from individual patients. Clinical data included demographic variables, comorbidities, operative details, postoperative outcomes, and intermediate-term follow-up results. To confirm preoperative atrial fibrillation, we reviewed the preoperative electrocardiogram. Coronary artery disease was defined as ≥50 % obstruction of a coronary vessel that may or may not have been amenable to bypass. Emergency operations were performed within 24 h after the clinical diagnosis. Carotid artery disease was defined as ≥50 % obstruction, per carotid artery duplex scan findings. Pulmonary hypertension was defined as right ventricular systolic pressure (RVSP) ≥40 mmHg, per echocardiography findings.

Endpoints

Endpoints were postoperative hemodynamic measurements, in-hospital morbidity and mortality, valve-related events, resource utilization, clinically significant structural valve degeneration and intermediate-term survival rates. To define in-hospital morbidity (perioperative myocardial infarction, respiratory failure, sepsis, renal failure, and neurologic complications), we used the Society of Thoracic Surgeons national Adult Cardiac Surgery database. For surviving patients, we used hospital length of stay as a surrogate marker for resource utilization. Clinically relevant structural valve degeneration was defined as the need for reoperation or severe regurgitation or stenosis preceding death.

Hemodynamic measurements

For the echocardiography findings, we retrieved the last available transthoracic study from our server for each patient; for all quantifications, we used Syngo software (Siemens AG, Munich, Germany), with interpretations made by two experienced echocardiologists (A.S and W.S.). For the 2D echocardiography analysis of LVOT diameter, left ventricular ejection fraction (EF), and stroke volume (SV), we used the biplane method of disks (modified Simpson’s rule). We also recorded RVSP, peak and mean transmitral gradients (mmHg), left atrial (LA) size, left ventricular internal diameter (LVID), and effective orifice area (EOA).

To calculate the LVOT systolic velocity time integral (VTI), we used the mean of three waveforms acquired by LVOT pulse-wave Doppler; then, to calculate SV, we multiplied LVOT diameter by LVOT VTI. If systolic LVOT VTI was not available, we obtained SV by using the biplane method of disks (modified Simpson’s rule).

To calculate the peak transmitral gradient, we obtained measurements from mitral diastolic continuous-wave Doppler assessment and then used the simplified Bernoulli equation. To calculate the mean transmitral gradients, we obtained measurements from the diastolic waveform VTI. For both the peak and mean measurements, we used the mean of three waveforms.

To calculate the mitral valve area (MVA), we used the pressure half-time (t1/2) method (MVA = 220/t1/2). To calculate the mitral EOA, we divided SV by the mitral diastolic VTI. To calculate the EOA index, we divided the EOA by body surface area.

To assess the presence and degree of mitral regurgitation, we used the four-chamber view per color flow Doppler imaging, according to the semiquantitative approach recommended by the American Society of Echocardiography. In six of the 70 patients, postoperative transthoracic 2D echocardiograms were not available for analysis.

Data analysis

Continuous demographic variables are expressed as the mean ± the standard error of the mean (SE); categorical values are expressed as the n (percentage). To compare pre- and postoperative echocardiography findings, we used the paired Student t test. To assess intermediate-term survival (in months) after Magna valve implantation, we used the Kaplan-Meier method, with censoring (N = 70, with n = 16 deaths, 54 censored). To assess freedom from structural valve degeneration we used the Kaplan-Meier method (N = 70, with n = 2 events, 68 censored). To analyze the relationship between dichotomous variables and survival, we used the 2-tailed Fisher exact test.

Results

Preoperative demographic and clinical characteristics are reported in Table 1. The mean age was 68 years (range, 29 to 88). Of the 70 patients, 30 (43 %) had New York Heart Association (NYHA) class III or IV disease. The Magna valve implantation was considered an emergency operation in 3 (4.3 %) of the patients; 9 (12.8 %) were in cardiogenic shock preoperatively. Additionally, 36 (51.4 %) of the patients had moderately severe, or worse, mitral regurgitation; 21 (30 %), severe mitral stenosis. In terms of previous surgery, 30 (42.8 %) of the patients had undergone at least 1 previous cardiac operation; 19 (27.1 %), a previous mitral valve repair or replacement.

Table 1 Patient characteristics (N = 70)a

Operative factors and findings are summarized in Table 2. Of the 70 patients, 69 (98.6 %) underwent a full sternotomy. The most common approach to Magna valve implantation was transeptal with subvalvular leaflet preservation. The most common Magna valve size was 29 mm, followed by 25 mm. Mitral annular calcification was the most common operative finding, followed by degenerative disease. Most patients had at least 1 additional procedure performed at the time of the Magna valve implantation.

Table 2 Operative factors and findings (N = 70)a

Thirty eight patients had at least some MAC documented in the operative note. Moderate to severe MAC was noted in 29 of these cases (57 %). MAC involved the anterior annulus in 22 cases (57 %) and the posterior annulus in 36 cases (94 %). In 17 cases (44 %) the MAC was left undisturbed.

Twelve cases (31 %) required at least some debridement and nine cases (23 %) required extensive debridement. Annular reconstruction with a patch was used to avoid AV rupture if adipose tissue was visible on the ventricular side of the annulus after debridement. Three cases (8 %) used a bovine pericardial patch and 1 case (3 %) used a felt patch. Only one case required an additional bypass run for mild-moderate perivalvular regurgitation to reinforce a separation along the anterior rim. The rest had either no or trivial MR at the conclusion of the replacement.

Postoperative complications are detailed in Table 3. We noted five in-hospital deaths, for a mortality rate of 7.4 % in the immediate postoperative period. For the remaining 65 survivors, the mean hospital length of stay was 12 days (range, 4 to 42). Two patients (2.9 %) suffered a stroke within 30 days. The most common rhythm disturbance was atrial fibrillation (71 %). In all, 6 (8.5 %) of the patients developed renal failure requiring dialysis; 3 (4.2 %) suffered an in-hospital cardiac arrest.

Table 3 Postoperative course and complications (N = 70)a

The average clinical follow-up was 16 months (range, 0–62 months). The 2- and 5-year survival rates were 84 and 40 % respectively. The median survival was 4.7 years (Fig. 2). Several dichotomous perioperative factors were associated with nonsurvival, including preoperative atrial fibrillation, ischemic mitral regurgitation, use of an intraaortic balloon pump (IABP), cardiogenic shock, cardiac arrest, and renal failure (Table 4). Interestingly, RVSP ≥40 mmHg was associated with a higher survival rate on univariate analysis (P = 0.02) suggesting, at minimum, that no penalty was incurred for Magna valve implantation in patients with moderate pulmonary hypertension.

Fig. 2
figure 2

Kaplan-Meier non-parametric estimate of all-cause mortality based on the observed survival in patients with the Carpentier-Edwards Magna valve in the mitral position with censoring (N = 70, with n = 16 deaths, 54 censored). X-axis represents time after valve implant, in months. aNumber at risk for each 10 month interval beginning with n of 70

Table 4 Factors associated with increased mortality (N = 70)a

Pre- and postoperative echocardiography findings are compared in Table 5. The mean follow-up time for postoperative echocardiography was 5 months (range, 0.3 to 50 months). Per their final echocardiogram during our study period, 98 % of surviving patients had trivial or no mitral regurgitation, one patient had mild mitral regurgitation and one patient had severe mitral regurgitation. No short term valve related events such as vegetations, dehiscence, degeneration, stenosis, or thrombosis were documented on any of the postoperative echocardiograms. However, one patient had obstruction of the LVOT by a strut requiring a reoperation at 2 years.

Table 5 Pre- and postoperative echocardiography findings (N = 70)a

Two patients had evidence of severe structural valve degeneration on follow-up (Fig. 3). In one, severe prosthetic MR was present at 2 years and the patient died before a reoperation. In the other patient, severe mitral calcification was noted during a redo AVR 6 months after the initial Magna valve implantation. The postoperative EF decreased by a mean of 2.5 % across matched pairs, although the difference was not statistically significant (Table 5). We observed a trend toward a reduction in LA size and in LVID.

Fig. 3
figure 3

Kaplan-Meier non-parametric estimate of freedom from structural valve degeneration, as defined by a need for reoperation due to valve dysfunction preceding death, in patients with the Carpentier-Edwards Magna valve in the mitral position (N = 70, with n = 2 events, 68 censored). X-axis represents time after valve implant, in months. aNumber at risk for each 10 month interval beginning with n of 70

The peak transmitral gradient decreased as the Magna valve size increased (Fig. 4). The same was true for the mean transmitral gradient. However, we noted no statistically significant differences in the mean gradient between the two most commonly employed valve sizes (29 and 25 mm), suggesting that even a small Magna valve was capable of yielding a low mean gradient. Although we noted a trend toward lower RVSP values for each Magna valve size, the only statistically significant decrease was with 31 mm (Fig. 5). The Effective Orifice Area Index (EOAI) did not vary significantly between valve sizes except for the 31 mm valve which had a statistically higher value than the 25 mm valve (Fig. 6).

Fig. 4
figure 4

Postoperative peak (gray circles) and mean (dark boxes) transmitral gradients (in millimeters of mercury) per echocardiography by Magna valve size (in millimeters). Asterisks = means that are significantly different (P < 0.05)

Fig. 5
figure 5

Preoperative (gray circles) and postoperative (dark boxes) right ventricular systolic pressure (RVSP) (in millimeters of mercury) per echocardiography by Magna valve size (in millimeters). Asterisks = pre- and postoperative matched pairs that are significantly different (P < 0.05)

Fig. 6
figure 6

Effective Orifice Area Index (EOAI) across various valve sizes. The asterisk denotes a significant difference between the 31 and 25 mm EOAI

Discussion

Our 5-year study is the first to objectively evaluate the hemodynamic and clinical outcomes with the Magna valve, a modification of the earlier Carpentier-Edwards Perimount mitral valve bioprosthesis. We found excellent intermediate-term durability with this versatile design, which is well suited for complex situations; even at the lowest valve sizes, this new model provided patients with excellent hemodynamics.

Most of the patients in our series had extensive comorbidities, a high NYHA class, severe valve disease, and previous cardiac operations, many of which involved the mitral valve. Thus, our patients were typical of those undergoing mitral valve replacement in the current era [1, 4, 13]. While the survival rate at 2 years of 84 % was excellent, the survival rate of 40 % at 5 years was commensurate with this cohorts’ extent of illness. For comparison, the St. Jude Medical Biocor valve series from Rizzoli et al. noted a 5-year survival rate of 54 % [12]. The 15 year experience with the original Perimount bioprosthesis showed a 5 year survival of 76 %, although their cohort had fewer comorbidities and less reoperations than observed in the current study [7]. We also observed significant postoperative morbidity, including renal failure, cardiogenic shock, and respiratory failure. One patient suffered a death due to AV groove disruption which was related to severe MAC rather than the Magna valve itself. One reoperation was required at 2 years due to LVOT obstruction from a strut. These cases highlight the need for caution even with this low-profile design. The freedom from structural valve degeneration of 90 % was excellent and compares favorably with that observed in other series [7, 12]. One reoperation was due to structural valve degeneration at 6 months and one patient died due to severe MR at 2 years without a reoperation.

We identified several factors associated with increased mortality in our patients, including atrial fibrillation, cardiogenic shock, IABP use, cardiac arrest, and ischemic mitral regurgitation. Wang et al. also found significantly increased mortality in patients with preoperative atrial fibrillation, but they found improved survival with a concomitant Cox maze and LA ligation [14]. In our series, despite the 27 % incidence of preoperative atrial fibrillation, only 10 % of our patients underwent an antiarrhythmic procedure—perhaps a reflection of the length and complexity of the procedures, along with the frailty of the tissues, which discouraged ancillary procedures that were not absolutely necessary in these older patients.

De Bonis et al. recently showed that ischemic mitral regurgitation in sicker patients was associated with lower survival when the mitral valve was replaced rather than repaired [15]. Conversely, Gillinov et al. found no difference in survival between repair and replacement in patients with a higher NYHA class and complex regurgitant jets [4]. This is consistent with the randomized controlled prospective trial of Acker and colleagues that showed no survival benefit for repair over replacement and greater durability with valve replacement [16]. We believe it is reasonable to use the Magna valve for ischemic mitral regurgitation, although repair should be considered in appropriate candidates.

What are the advantages of a lower-profile bioprosthesis? Patients receiving a mitral prosthesis have a natural reduction in the LVOT [17]. In addition, the struts of a bulky bioprosthetic can protrude into the LVOT, causing a clinically significant increase in gradients [18]. A smaller, lower-profile valve reduces such concerns. Reoperative or multiple-valve procedures in patients with mitral annular calcification or degenerative disease pose additional restrictions on the prosthesis size. The exposure in such procedures is challenging; the mitral annulus and left ventricular cavities are often restricted. Most patients in our series underwent a full sternotomy, with transseptal mitral exposure and with implantation of a small Magna valve (25 to 29 mm). In addition, in patients with ischemic mitral regurgitation, the Magna valve’s low-profile design is more likely to allow preservation of the subvalvular apparatus which may be associated with improved survival [19, 20]. In our series, 80 % of patients had preservation of some portion of the subvalvular apparatus, with no reported outflow obstruction.

The mean gradients achieved with the Magna valve in our series compared favorably with those reported in previous publications on other mitral bioprostheses [1012]. Our lowest gradients were achieved with the 27 and 31-mm Magna valves, although the gradients achieved with the 25-mm Magna valve were only slightly higher. Thus, the 25-mm valve size is a reasonable option for patients with a constrained annulus or small ventricle, especially when left ventricular outflow tract obstruction (LVOTO) is a concern. While the low profile design may reduce the incidence of LVOTO, care must still be taken to keep the struts out of the way of the LVOT as one case in this series did demonstrate LVOTO resulting in redo-replacement.

The valve size of 25 was the second most common size used in this series which suggests small working annuluses. This could be explained by variations in patient anatomy or surgical technique, high rate of MAC with limited debridement, and/or the presence of reoperations. We tend to use either inverting (ventricular to atrial) or everting (atrial to ventricular) suture techniques depending on anatomy and ease of placement. We favor inverting sutures in the setting of MAC which places the sewing ring on the atrial side of the mitral annulus. If everting sutures are used then the valve may sit slightly intra-annular. We will infrequently construct a neo-inner annulus as described by Di Stefano and colleagues to avoid debridement in the setting of severe MAC [21]. The presence of favorable mean gradients suggests that the range of Magna sizes can accommodate the operator’s preference and/or patient’s anatomy.

The favorable hemodynamic profile of the Magna valve is further supported by the significant reduction we observed in RVSP. It is anticipated that a persistent reduction in RVSP could correlate with improved long-term survival, but we did not have a large enough sample size or long enough follow-up to definitively show any correlation [22]. The suggestion that even smaller Magna valve sizes produce favorable hemodynamics is supported by the fact that we saw no appreciable difference in EOA index across valve sizes, except for the 31-mm valve.

Our study supports the use of the Magna valve as a replacement option, but we acknowledge several important limitations. First, echocardiographic follow-up was not as standardized as it is for repairs and not all patients had echos at their longest follow-up interval. Echocardiography was generally obtained if a clinical indication was met. Thus, our review of structural valve degeneration was limited to the causes of death, reoperation and the last available echo prior to these events. The Magna valve design is based on the previous Perimount model, which is associated with 14 years’ worth of data showing freedom from explantation and from structural valve deterioration, so we are confident that the newer model will do at least as well. Additionally, the prospective evaluation of the echocardiograms by two experienced cardiologists was crucial as it standardized the valve’s performance measures. But it was limited to a mean follow-up of 5 months with a broad range of intervals. We elected to evaluate the last echos to avoid an unfair advantage from early postoperative performance. We noted a high interobserver variability for the calculation of the EOA index in our series, making it difficult to reach conclusions about the actual value for any given Magna size. However, we were able to make statistical comparisons for changes in the EOA index across valve sizes. Our study was underpowered for a multivariate analysis, although trends were established by univariate measures.

The peak gradient for the 25, 29 and 31 mm valves gradually decreased as expected. However, the peak was higher for the 29 mm valve than the 27 mm valve. Similarly, the 25, 27 and 31 mm mean gradients gradually decreased as expected. But the mean for the 29 and 25 mm valves were not statistically different. Peak and mean gradients can be affected by various transient factors such as cardiac output, heart rate, and volume status. In addition, the patient population was heterogeneous and we did not have an equal distribution of patients throughout the various valve sizes. Thus, we can conclude that the peaks and means decreased in our series with larger valve sizes as expected but we cannot give definitive estimates of these relative gradients that would translate to a broader population.

Conclusions

In conclusion, this is the first published report on outcomes after implantation of the Magna valve. Modifications are regularly made to valves; it is important that at least short-term results are made available. In our 5-year study, we found excellent freedom from structural valve degeneration, excellent short term survival, good hemodynamics at even the lowest valve sizes, substantial echocardiographic improvement, and minimal valve-related events. The Magna valve’s low-profile design and ThermaFix process make it ideal for patients who have small ventricles, extensive comorbidities, or complex valve disease, as well as for those who need reoperations and concomitant cardiac procedures, including valve replacement.

References

  1. Gammie JS, Sheng S, Griffith BP, Peterson ED, Rankin JS, O’Brien SM, et al. Trends in mitral valve surgery in the United States: results from the Society of Thoracic Surgeons Adult Cardiac Surgery Database. Ann Thorac Surg. 2009;87(5):1431–7. discussion 7–9.

    Article  PubMed  Google Scholar 

  2. Bonow RO, Carabello BA, Chatterjee K, De Leon Jr AC, Faxon DP, Freed MD, et al. 2008 Focused update incorporated into the ACC/AHA 2006 guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Revise the 1998 Guidelines for the Management of Patients With Valvular Heart Disease): endorsed by the Society of Cardiovascular Anesthesiologists, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. Circulation. 2008;118(15):e523–661.

    Article  PubMed  Google Scholar 

  3. Gillinov AM, Blackstone EH, Nowicki ER, Slisatkorn W, Al-Dossari G, Johnston DR, et al. Valve repair versus valve replacement for degenerative mitral valve disease. J Thorac Cardiovasc Surg. 2008;135(4):885–93. 93 e1-2.

    Article  PubMed  Google Scholar 

  4. Gillinov AM, Faber C, Houghtaling PL, Blackstone EH, Lam BK, Diaz R, et al. Repair versus replacement for degenerative mitral valve disease with coexisting ischemic heart disease. J Thorac Cardiovasc Surg. 2003;125(6):1350–62.

    Article  PubMed  Google Scholar 

  5. Grossi EA, Galloway AC, Zakow PK, Miller JS, Buttenheim PM, Baumann FG, et al. Choice of mitral prosthesis in the elderly. An analysis of actual outcome. Circulation. 1998;98(19 Suppl):II116–9.

    CAS  PubMed  Google Scholar 

  6. Kulik A, Bedard P, Lam BK, Rubens FD, Hendry PJ, Masters RG, et al. Mechanical versus bioprosthetic valve replacement in middle-aged patients. Eur J Cardiothorac Surg. 2006;30(3):485–91.

    Article  PubMed  Google Scholar 

  7. Marchand MA, Aupart MR, Norton R, Goldsmith IR, Pelletier LC, Pellerin M, et al. Fifteen-year experience with the mitral Carpentier-Edwards PERIMOUNT pericardial bioprosthesis. Ann Thorac Surg. 2001;71(5 Suppl):S236–9.

    Article  CAS  PubMed  Google Scholar 

  8. Ruel M, Chan V, Bedard P, Kulik A, Ressler L, Lam BK, et al. Very long-term survival implications of heart valve replacement with tissue versus mechanical prostheses in adults <60 years of age. Circulation. 2007;116(11 Suppl):I294–300.

    PubMed  Google Scholar 

  9. Eichinger WB, Hettich IM, Ruzicka DJ, Holper K, Schricker C, Bleiziffer S, et al. Twenty-year experience with the St. Jude medical Biocor bioprosthesis in the aortic position. Ann Thorac Surg. 2008;86(4):1204–10.

    Article  PubMed  Google Scholar 

  10. Firstenberg MS, Morehead AJ, Thomas JD, Smedira NG, Cosgrove 3rd DM, Marchand MA. Short-term hemodynamic performance of the mitral Carpentier-Edwards PERIMOUNT pericardial valve. Carpentier-Edwards PERIMOUNT Investigators. Ann Thorac Surg. 2001;71(5 Suppl):S285–8.

    Article  CAS  PubMed  Google Scholar 

  11. Goetze S, Brechtken J, Agler DA, Thomas JD, Sabik 3rd JF, Jaber WA. In vivo short-term Doppler hemodynamic profiles of 189 Carpentier-Edwards Perimount pericardial bioprosthetic valves in the mitral position. J Am Soc Echocardiogr. 2004;17(9):981–7.

    Article  PubMed  Google Scholar 

  12. Rizzoli G, Bottio T, Vida V, Nesseris G, Caprili L, Thiene G, et al. Intermediate results of isolated mitral valve replacement with a Biocor porcine valve. J Thorac Cardiovasc Surg. 2005;129(2):322–9.

    Article  PubMed  Google Scholar 

  13. Zhou YX, Leobon B, Berthoumieu P, Roux D, Glock Y, Mei YQ, et al. Long-term outcomes following repair or replacement in degenerative mitral valve disease. Thorac Cardiovasc Surg. 2010;58(7):415–21.

    Article  CAS  PubMed  Google Scholar 

  14. Wang B, Xu ZY, Han L, Zhang GX, Lu FL, Song ZG. Impact of preoperative atrial fibrillation on mortality and cardiovascular outcomes of mechanical mitral valve replacement for rheumatic mitral valve disease. Eur J Cardiothorac Surg. 2013;43(3):513–9.

    Article  PubMed  Google Scholar 

  15. De Bonis M, Ferrara D, Taramasso M, Calabrese MC, Verzini A, Buzzatti N, et al. Mitral replacement or repair for functional mitral regurgitation in dilated and ischemic cardiomyopathy: is it really the same? Ann Thorac Surg. 2012;94(1):44–51.

    Article  PubMed  Google Scholar 

  16. Acker MA, Parides MK, Perrault LP, Moskowitz AJ, Gelijns AC, Voisine P, et al. Mitral-valve repair versus replacement for severe ischemic mitral regurgitation. N Engl J Med. 2014;370(1):23–32.

    Article  PubMed Central  CAS  PubMed  Google Scholar 

  17. Rosendal C, Hien MD, Bruckner T, Martin EO, Szabo G, Rauch H. Left ventricular outflow tract: intraoperative measurement and changes caused by mitral valve surgery. J Am Soc Echocardiogr. 2012;25(2):166–72.

    Article  PubMed  Google Scholar 

  18. Ducas RA, Jassal DS, Zieroth SR, Kirkpatrick ID, Freed DH. Left ventricular outflow tract obstruction by a bioprosthetic mitral valve: diagnosis by cardiac computed tomography. J Thorac Imaging. 2009;24(2):132–5.

    Article  PubMed  Google Scholar 

  19. Borger MA, Yau TM, Rao V, Scully HE, David TE. Reoperative mitral valve replacement: importance of preservation of the subvalvular apparatus. Ann Thorac Surg. 2002;74(5):1482–7.

    Article  PubMed  Google Scholar 

  20. Reardon MJ, David TE. Mitral valve replacement with preservation of the subvalvular apparatus. Curr Opin Cardiol. 1999;14(2):104–10.

    Article  CAS  PubMed  Google Scholar 

  21. Di Stefano S, Lopez J, Florez S, Rey J, Arevalo A, San RA. Building a new annulus: a technique for mitral valve replacement in heavily calcified annulus. Ann Thorac Surg. 2009;87(5):1625–7.

    Article  PubMed  Google Scholar 

  22. Roselli EE, Abdel Azim A, Houghtaling PL, Jaber WA, Blackstone EH. Pulmonary hypertension is associated with worse early and late outcomes after aortic valve replacement: implications for transcatheter aortic valve replacement. J Thorac Cardiovasc Surg. 2012;144(5):1067–74. e2.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Gabriel Loor.

Additional information

Competing interests

The authors declare that they have no competing interests.

Authors’ contributions

GL carried out the study design, chart review, figure design and study oversight. KM contributed to study concept, design and sequence alignment. AS and WS read the echos and interpreted them in a prospective fashion. JD assisted with organization of data tables and results sequence alignment. VC performed the statistics. AR contributed to the study design and organization of data tables. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Loor, G., Schuster, A., Cruz, V. et al. The Carpentier-Edwards Perimount Magna mitral valve bioprosthesis: intermediate-term efficacy and durability. J Cardiothorac Surg 11, 20 (2016). https://doi.org/10.1186/s13019-016-0412-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-016-0412-4

Keywords