- Case report
- Open access
- Published:
An alternative solution for patient with high risk of coronary obstruction underwent TAVI procedure using a novel second-generation device – a case series
Journal of Cardiothoracic Surgery volume 14, Article number: 47 (2019)
Abstract
Background
Obstruction of the left or right coronary artery is a rare but lethal complication during transcatheter aortic valve implantation (TAVI). The new J-Valve™ prosthesis is a new second generation TAVI device which has several features to avoid the coronary obstruction such as low profile design and clip fixation of the native leaflets. The aim of this study is to report our initial experience of using this valve in treating patient with high risk factors for coronary obstruction during TAVI procedure.
Case presentation
Three high surgical risk patients (All females with 77, 76, and 75 years old) with symptomatic aortic stenosis were enrolled. All patients have the common feature of low coronary ostium height (< 10 mm) with narrowed aortic sinus (< 30 mm) on CT angiogram and marked leaflet calcification. Three 25 mm J-Valve prostheses were successfully implanted through trans-apical approach. No coronary obstruction was noted for these patients. Effective aortic open area was significantly increased after valve implantation (Preoperative 0.7, 0.7 and 0.65 cm2 – Postoperative 1.8, 1.9 and 2.0 cm2). Only one patient was noted to have trivial degree paravalvular leakage.
Conclusion
The new J-Valve prosthesis is a new second generation TAVI device. This system may provide another safety treatment option for patient with high risk factor for coronary obstruction underwent TAVI procedure.
Background
Transcatheter aortic valve implantation (TAVI) has evolved as a routine procedure to treat selected high-risk patients with severe aortic stenosis. Although obstruction of the left or right coronary ostium is a relatively rare but lethal complication during TAVI procedure, it is generally accepted that a distance between the coronary ostium and the annulus less than 10 mm or sinus of Valsalva diameter below 30 mm, and existing of bulky calcified leaflet are risks for coronary occlusion as for several type of classic TAVI devices such as Edward SAPIEN and Medtronic CoreValve [1, 2]. The J-Valve™ system is a brand new second-generation TAVI device featured by three U-shape anatomically oriented devices “claspers” which could facilitate intuitive “self-positioning” valve implantation [3]. Due to the ability of embracing the native calcified leaflets during valve implantation and specially designed bear area on valve stent, this low-profile designed valve is especially suitable for patient who is at high risk for coronary obstruction. We then reported our initial experience of using J-Valve in treating patient with high risk factor for coronary obstruction during TAVI procedure.
Case presentation
Three high surgical risk patients with symptomatic aortic stenosis who are at high risk for coronary occlusion during TAVI procedure were enrolled (three female patients). The basic characters shown in Table 1. These three patients have the common feature of low coronary ostium height (< 10 mm) with narrowed aortic sinus (< 30 mm). The effective aortic valve opening area (EOA) was 0.7 cm2, 0.7 cm2 and 0.65 cm2 in each patient with marked leaflet calcification. Aortic valve annular size was 24 mm, 25 mm, 24 mm on CT image.
The TAVI procedure using J-Valve™ prosthesis was performed through trans-apical approach (Fig. 1). The procedure detail is described previously [3]. Briefly the apical puncture was done with a super-stiff guide-wire placed into the descending aorta. A 14F sheath was carefully introduced into the left ventricle followed by a balloon catheter. Pre-procedural balloon-valvuloplasty of the native valve was performed under rapid pacing. In parallel, a J-Valve™ prosthesis was crimped into the 27-F Ausper-AS delivery system (Three 25 mm prostheses were used). The delivery system was bluntly inserted into the left ventricle and advanced under fluoroscopic guidance into a supra-annular position. Three ‘U-shaped’ graspers were then completely released first and carefully placed into the corresponding aortic sinus by gently pulling back the delivery system, thereby embracing the native leaflets. The angiogram was then performed to confirm that all the graspers were positioned correctly into the each aortic sinus. The valve was retrieved back gently into the annular plan with the guidance of the graspers and deployed without rapid ventricular pacing. After wound closure, the patient was transferred to the intensive care unit. Further postoperative course was uneventful. At discharge, TTE revealed an excellent valve function with EOA (1.8 cm2, 1.9 cm2, 2.0 cm2 in each patient) without any relevant incompetence. One patient was noted to have trivial degree paravalvular leakage.
Discussion
Although the incidence of coronary obstruction reveled in large TAVI series and registries are relatively low, it is still the most lethal complication of TAVI procedure [1]. Evidences have shown that a distance between the coronary ostium and the annulus less than 10 mm or sinus of Valsalva diameter below 30 mm, and existing of bulky calcified leaflet are risks for coronary occlusion as for classic TAVI device such as Edward SAPIEN and Medtronic CoreValve due to their design [2]. Contrast to western population, Asian patients always associated with low body weight, relatively small aortic root and low coronary orifice height. In some specific situation, choose of the classic TAVI device may warrant caution or even contraindicated.
The J-Valve™ system is a new second-generation self-expendable low-profile prosthesis that is featured by three U-shape anatomical orientated devices — “Clasper” encircling around the valve frame. The design concept of this device is to facilitate intuitive ‘self-positioning’ valve implantation through embracing the native aortic valve leaflet. In this case serious, successful TAVI procedure was done using this unique system in all three patients who are at high risk for coronary obstruction with low coronary height and narrowed aortic sinus. This valve have several unique features to minimized the risk coronary obstruction including 1) More accurate position of the device with anatomical orientated devices to avoid malposition; 2) Low-profile design with extra bare mental area in coronary orifice region (Fig. 2); 3) Unlike the traditional design of classic stent valve system, the unique designed U-shape claspers could embrace the native aortic valve leaflet (clip mechanism) then subsequently leaving the coronary ostia freely accessible from heavily calcified aortic leaflet; 4) Meanwhile, J-Valve™ system has the advantage of achieving better anchoring through embrace the leaflets, then diminish the necessity of oversizing and minimize the risk of coronary obstruction.
Conclusion
The new J-Valve™ prosthesis is a brand new second generation TAVI device. This system may provide a safety treatment option for patient with high risk factor for coronary obstruction underwent TAVI procedure.
Abbreviations
- EOA:
-
Effective aortic valve opening area
- TAVI:
-
Transcatheter aortic valve implantation
References
Ribeiro HB, Webb JG, Makkar RR, et al. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. JACC Cardiovasc Interv. 2013;6:452–61.
Ribeiro HB, Nombela-Franco L, Urena M, et al. Coronary obstruction following transcatheter aortic valve implantation: a systematic review. JACC Cardiovasc Interv. 2013;6:452–61.
Zhu D, Wei L, Cheung A, et al. Treatment of Pure Aortic Regurgitation Using a Second-Generation Transcatheter Aortic Valve Implantation System. J Am Coll Cardiol. 2016;67:2803–5.
Acknowledgements
Not applicable.
Funding
None.
Availability of data and materials
The dataset of this case report is available from the corresponding author on reasonable request.
Author information
Authors and Affiliations
Contributions
HQ performed the statistical analysis and drafted the manuscript. YC participated in the surgery procedure and performed the CT scan. ZC participated in the design of the study and performed the statistical analysis. YG conceived of the study, and participated in its design and coordination and helped to draft the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
The hospital ethics committee approval was granted of this case report.
Consent for publication
Written informed consent was obtained from the patient for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.
Competing interests
The authors declare that they have no competing interests.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Qian, H., Chen, Y., Cheng, Z. et al. An alternative solution for patient with high risk of coronary obstruction underwent TAVI procedure using a novel second-generation device – a case series. J Cardiothorac Surg 14, 47 (2019). https://doi.org/10.1186/s13019-019-0859-1
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s13019-019-0859-1