Early and midterm outcomes of a bentall operation using an all-biological valved BioConduit™

Abstract

Objectives

To analyze the midterm results of aortic root replacement using the valved, all biological, No React®, BioConduit™.

Methods

From 2017 to 2020, we prospectively followed 91 consecutive patients who underwent a Bentall procedure with a BioConduit™ valved graft in our institution. The primary outcomes were aortic bioprosthetic valve dysfunction and mortality according to Valve Academic Research Consortium 3 (VARC3).

Results

Mean age was 70 ± 10 years and 67 patients (74%) were men. Ascending aortic aneurysm (72%), aortic valve regurgitation (51%) or stenosis (20%) and acute endocarditis (14%) were the main indications for surgery. Seventy-four patients (81.3%) were followed up at 1 year. The perioperative mortality was 8% (n = 8), the early, 1 year, mortality was 2% (n = 2) and the midterm mortality, at 4 years of follow up, was 4% (n = 3). Ten patients fulfilled the criteria for hemodynamic valve deterioration at 1 year (13%) and 14 for a bioprosthetic valve failure during the entire follow-up (17%).

Conclusions

We are reporting early and midterm results of Bentall procedures with the all-biological, valved, No-React® BioConduit™. To our knowledge, this is the first study reporting an early and midterm unexpectedly high rate of non-structural prosthetic hemodynamic deterioration. The rate of endocarditis and atrioventricular disconnections remain similar to previous studies.

Peer Review reports

Background

The Bentall procedure is the gold standard therapy in patients with either ascending aorta or aortic root aneurysm combined with aortic valve disease precluding a valve sparing procedure. [1]. The original technique described by Bentall and De Bono using a composite mechanical valved graft benefited from iterative refinements in order to overcome specific surgical drawbacks [2]. Nowadays, either preassembled or self-assembled conduits, associating tubular straight or Valsalva graft and biological or mechanical valve, are widely used [3, 4]. As an alternative, a fully xenobiological stentless valved conduit, the Shelhigh NR-2000, was introduced in the late 1990’s and thereafter withdrawn from the market. Recently, a totally biological stentless conduits have been reintroduced in a modified version, using a porcine aortic valve and a bovine pericardial tube (BioIntegral™, BioValsalva™). The goal of our study was to investigate the early and midterm results of the Bentall procedure using BioIntegral™ BioConduit™ in our single-center experience.

Methods

Study design

It was a prospective observational study, without control-group, carried out from 2017 to 2020 in our tertiary centre.

Study population

All consecutive patients undergoing a Bentall procedure with the BioIntegral™ Surgical BioConduit™ No-React® in our institution (n = 91) were prospectively included during the study period. All patient data were collected from hospitals’ medical records. Cohort patients underwent an aortic root replacement with an all-biological graft in cases of complex endocarditis or redo surgeries, cases of patients who were not candidates for an autograft procedure but a mechanical graft was not indicated and in elderly patients with degenerative aortic root disease.

The study is conformed to the principles outlined in the Declaration of Helsinki. According to French law on ethics, patients were informed that their codified data would be used for the study. The patient also provided informed written consent for the publication of the study data. The Institutional Review Board of the Rangueil University Hospital of Toulouse, France, approved the study protocol and the publication of data (number RnlPH 2022-50) on the 7th of April 2022.

Surgical data

All interventions were performed by four senior surgeons. All patients were operated through median sternotomy. Standard cardiopulmonary bypass, aortic cross-clamp and anterograde and retrograde cold blood potassium-enriched cardioplegia were used in all patients. When an aortic arch replacement was performed, moderate hypothermic circulatory arrest and selective antegrade cerebral perfusion through the right axillary artery were associated. After aortic cross-clamping, the aortic valve, the aortic root and the ascending aorta were excised, followed by aortic annulus decalcification when required. The coronary ostia were isolated with their buttons. After sizing of the aortic annulus, the prosthesis was chosen, with a trend in oversizing. The all-biological stentless valved bioconduit is designed to offer similar hemodynamics as the native aortic valve and theoretically aimed to insure a larger post-operative effective orifice area (EOA). The conduits’ No-React® treatment aims to reduce the grafts’ related infections and calcifications as well as to prevent remodeling and graft aneurysmal dilatation [5,6,7]. As the traditional glutaraldehyde preserved biological valves tend to calcify, the No-React® detoxification process promises to eliminate residual glutaraldehyde and to ensure stable tissue cross-linking, resulting in less or no calcification or tissue deterioration in the animal model [5, 6]. A variety of BioConduit™ sizes between 21 to 29 mm were implanted. The trend of oversizing refers to choosing a conduit one-size oversized following the manufacturer’s recommendations, to achieve greater effective surface areas, avoiding mismatch [7]. Depending on operator preferences, two types of implantation techniques were used, either multiple single, interrupted, non-everting, reinforced U-stiches to implant the composite graft in supra-annular position (n = 33, 36%) or everting stitches to implant the prosthesis in intra-annular position (n = 58, 64%). Then, 2 holes were made in the tubular graft and both coronary ostia were reimplanted by running sutures (Prolene 6–0). Fibrin glue was used in most cases to reinforce sutures.

Imaging protocol

A post-operative transthoracic echocardiographic (TTE) assessment was performed before hospital discharge: 2D TTE standard views were obtained using a standard ultrasound system using a 1–5 MHz probe (VIVID S70, GE Healthcare). A new TTE assessment was conducted at 1 year, using the same system. EOA was calculated by the continuity equation method. Aortic annulus diameter was measured at mid-systole, from the parasternal long-axis view, at the level of the prosthetic annulus, in a zoomed mode, from inner-edge to inner-edge. The velocity–time integral of blood flow was measured in the left ventricular outflow track by pulsed doppler. Mean transaortic gradient and maximal velocity were evaluated by transprosthetic continuous wave doppler. The doppler velocity index was calculated as the ratio of the proximal peak flow velocity in the LVOT to the transprothetic peak flow velocity. All examinations were interpreted blindly on a dedicated workstation (EchoPac 204 GE Healthcare) by two operators. In addition, during the follow-up, in case of bioprosthetic valve failure (BVF), a TEE and a cardiac CT were performed.

Study outcomes

The endpoints used were those proposed by the Valve Academic Research Consortium 3 (VARC3).

Mortality [8]

  • Periprocedural mortality was defined as all-cause mortality occurring ≤ 30 days after the index procedure or occurring > 30 days but during the index hospitalization.

  • Early mortality was defined as death occurring > 30 days but ≤ 1 year after the index hospitalization.

  • Midterm mortality was defined as death occurring > 1 year after the index hospitalization but ≤ 4 years, at end of follow up.

Aortic bioprosthetic valve dysfunction (BVD)

  • In terms of etiology, BVDs were defined as structural valve deterioration (SVD), nonstructural valve dysfunction (NSVD), thrombosis or endocarditis. SVD was reported as intrinsic permanent changes of the prosthetic valve [9]. NSVD was reported as any abnormality not intrinsic to the bioprosthesis, resulting in its malfunction (e.g. pannus, prosthesis-patient mismatch) [9].

  • Hemodynamic valve performance assessment was protocolized at 1 year of the index procedure. Moderate hemodynamic valvular deterioration (HVD) was defined as an increase in the transaortic mean gradient of \(\ge\) 10 mmHg resulting in a mean gradient of \(\ge\) 20 mmHg, with concomitant decrease in the EOA \(\ge\) 0.3 cm2 and decrease in doppler velocity index \(\ge\) 0.1 compared to the post-operative assessment. Severe HVD was defined as an increase in the mean gradient of \(\ge\) 20 mmHg resulting in a mean gradient of \(\ge\) 30 mmHg, with concomitant decrease in the EOA \(\ge\) 0.6 cm2 and decrease in doppler velocity index \(\ge\) 0.2 [9].

  • During the follow-up, the occurrence of a BVF was considered as an endpoint. Finally, BVF was defined by the occurrence of BVD associated with clinically expressive criteria (heart failure symptoms, fever, angina, ischemic event), irreversible severe HVD, aortic valve reoperation or re-intervention or valve-related death [9].

Statistical analysis

Continuous variables were expressed as means ± standard deviation or as medians with interquartile ranges (IQR) when not normally distributed. Nominal variables were expressed as numbers and percentages. The association between the mean values of continuous variables was assessed using the Mann–Whitney rank sum test. Nominal variables were investigated by the χ2 test or the Fisher exact test when appropriate. The software XLSTATS v2019.1 (Addinsoft, Paris, FR) was used for statistical analysis.

Results

Population

This study included 91 patients, mostly men (74%) with a mean age at intervention of 70 ± 10 years. Preoperative patient’s characteristics are reported in Table 1. Most patients had an ascending aortic aneurysm (n = 65, 72%). There were 22 cases of redo procedures (24%). Sixteen procedures were performed on an emergency basis (18%), including 10 cases of type A acute aortic dissection (11%), 12 cases of severe prosthetic endocarditis (13%) and 1 complex of native aorto-mitral endocarditis. Forty-four patients underwent combined interventions (48%): coronary artery bypass graft in 22 cases (24%), mitral valve repair or replacement in 5 (6%) and aortic arch or hemiarch replacement in 14 (15.3%) cases.

Table 1 Patients and intra-operative characteristics (n = 91)

Among the 83 patients discharged from hospital, 9 were lost to follow-up (10%). Median follow-up was 4 years.

Mortality

The periprocedural mortality rate was 8% (n = 8). Among them, 3 patients with a pre-operative severe left ventricular dysfunction died shortly after the procedure from low cardiac output and multiorgan failure syndrome, 2 patients operated for complicated infective endocarditis on previous Bentall prosthesis died in the first 24 h from refractory septic vasoplegic syndrome, 2 patients operated for acute type A aortic dissection died either from massive hemorrhagic stroke or acute right ventricular failure in the first week after surgery and 1 patient experienced acute respiratory distress syndrome.

Early mortality rate was 2%. Among the 83 patients discharged from hospital, 2 died in the first year, both of them experiencing graft proximal anastomosis partial detachment at 9 and 10 months, respectively. Both were reoperated, but died from hemorrhagic stroke or massive intraoperative bleeding, respectively.

Midterm mortality rate was 4% (3 patients). One 62 years-old male, who had already undergone two Bentall procedures presented a recurrent bacterial graft endocarditis at 18 months after discharge. He was referred to surgery and died in the operating room from uncontrolled bleeding. One 83 years-old patient died from respiratory distress related to a severe form of Covid-19 pneumonia. The third, 72 years-old male died from metastatic pulmonary adenocarcinoma.

Bioprosthetic valve dysfunction

Early results

Regarding the prosthetic hemodynamic features, despite normal postoperative hemodynamic profiles, without signs of obstruction, we observed a decrease in graft performance at 1 year, mainly in terms of EOA (1.3 ± 0.2 cm2 vs 0.9 ± 0.4 cm2, p = 0.02 for 23 mm graft, 1.6 ± 0.4 cm2 vs 1.2 ± 0.5 cm2, p = 0.01 for 25 mm graft, 1.9 ± 0.5 cm2 vs 1.5 ± 0.5 cm2, p < 0.01 for 27 mm graft, 2.1 ± 0.6 cm2 vs 1.6 ± 0.6 cm2, p = 0.17 for 29 mm graft) but also in terms of transaortic mean gradient and maximum velocity. [10, 11].

At the 1-year follow-up, 10 patients fulfilled criteria for HVD (13%), 6 being moderate and 4 severe. The factors associated with HVD are presented in Table 2. Preoperative characteristics were not associated with HVD. The implanted graft caliber was significantly associated with the occurrence of HVD at 1 year (8 (80%) vs. 21 (33%), p = 0.01 for 23- or 25-mm graft), especially in the smallest sizes.

Table 2 Factors associated with the presence of hemodynamic valve deterioration (HVD) at 1 year (n = 74)

However, the patients who secondarily developed HVD had similar hemodynamic parameters as the rest of the cohort at the pre-discharge postoperative exam.

There was 1 case (1.3%) of early graft endocarditis with negative blood cultures, diagnosed at 8 months after surgery. We treated it medically.

We noted three cases (4.1%) of partial proximal anastomosis disruption: two cases with total atrio-ventricular disruption with large false aneurysms at the level of the proximal anastomosis at 9 months and respectively at 10 months after the initial surgery. We did not find any argument in favor of an infective endocarditis. Moreover, one patient’s initial surgery was in an elective setting, for a degenerative aortic aneurysm. The third patient had a similar aortic root disruption, but it appeared 4 months after a Bentall intervention for a complex aortic and mitral endocarditis with fragile tissues.

Midterm results

Five patients experienced infective endocarditis (6.8%) at a median of 1 year and 4 months. One patient was treated surgically and the other 5, medically.

There weren’t observed any more cases of hemodynamic dysfunction or AV disruptions until follow up was closed.

Regarding early and midterm results, we observed a total of 14 patients presenting a BVF during follow-up (17%). A re-intervention was performed in 8 cases (10%). Three patients had a valve-in-valve TAVR for severe HVD. Five patients had open redo surgery: three cases of partial proximal anastomosis detachment, one case of graft endocarditis with valvular involvement and one severe HVD. Among them, 3 patients died related to the procedure.

Regarding etiologies of BVF, 6 patients experienced infective endocarditis (8%), 3 patients had partial proximal anastomosis detachment (4%) and the other 4 patients had severe irreversible HVD (5.5%). All patients with endocarditis underwent TTE and TEE which were abnormal in two cases, highlighting valve involvement (aortic vegetations (n = 2), aortomitral abscess (n = 1) and pseudoaneurysm (n = 1)). Endocarditis patients also underwent a CT scan which showed in all cases a proximal collection around the biological graft with peripheral contrast diffusion, without signs of a pseudoaneurysm. PET combined with CT (PET/CT) was also performed in four cases, showing an abnormal intense uptake on the graft collection and/or the valvular prosthesis. Regarding the pathogens involved, there were 2 cases of staphylococcus involvement (aureus and epidermidis), 1 with Enterobacter Aerogenes, 1 with E. coli, another case a streptococcus oralis infection and lastly, 1 with negative blood cultures.

For all patients with HVD, thrombosis was ruled out by CT, and endocarditis by Duke criteria. SVD was eliminated by TEE and CT. Of note, none of these patients had regurgitation. Regarding NSVD, a patients-prosthesis mismatch did not appear to be involved, as postoperative indexed EOA was not associated with the occurrence of HVD at 1 year. Finally, for all patients with HVD, imaging found a 3–4 mm, circumferential, hyperechogenic ring, located at the level of the prosthetic aortic annulus and at the graft’s proximal anastomosis towards its’ ventricular side, near the pledgets. In these patients, on CT and TEE, there was no structural prosthetic abnormality (neither fibrosis, calcification, leaflet tear and wear, hypo-attenuated leaflet thickening nor thrombosis). Moreover, we have randomly reviewed patients’ echocardiographic exams concluding that the circumferential structure is a common finding in patients being implanted with this type of biological graft. Indeed, 29 patients had a reduction of their aortic annulus size at 1 year (39%). We identified this annular structure in the per-operatory setting, during the reintervention for a failing bioprosthesis (Fig. 1).

Fig. 1
figure 1

A Transesophageal echocardiography, deep transgastric view, annular hyperechogenic structure (yellow arrow); LVOT—left ventricle outflow tract; B Intraoperative view with the annular diaphragmatic structure (black arrow) visualized after complete prosthetic dissection

Discussion

This prospective study reports the midterm outcomes of 91 consecutive patients who underwent a Bentall procedure with the BioConduit™ No React® between 2017 and 2020, in our institution. Main results are as follows: 1) the hemodynamic performances at 1 year were unsatisfying, with an overall trend towards reduction in EOA and a significant rate of HVD, 2) despite the No-react® treatment of the conduit, graft endocarditis was not rare and 3) we observed some cases of early graft detachment at its proximal anastomosis with the aortic root.

Biological bioconduits, by avoiding stent and sewing cuff at the annular level, are intended to achieve more physiological flow pattern and thus superior hemodynamics [7, 12, 13]. However, our study identified a high rate of HVD and a general decrease in EOA at 1 year as being related to prosthesis of smaller calibers (23 and 25 mm) (p < 0.001) and renal insufficiency (p < 0.001) (Table 2). Even though we observed a general reduction in the annulus diameter, we think it had a more rapid obstructive impact in initially smaller calibers.

Regarding patients with HVD, our imaging protocol allowed us to rule out endocarditis, thrombosis or SVD in all cases. Our main observation explaining this phenomenon causing NSVD might be similar to a pannus formation. Firstly, we noticed a reduction in the aortic annulus diameter at 1 year (Fig. 2). In addition, we found a 3–4 mm, external circumferential, hyperechogenic, non-perfused ring, situated at the level of the prosthetic aortic annulus and at the graft’s proximal anastomosis near the pledgets. This structure was identified during our imaging protocol and confirmed in the perioperative setting (Fig. 1). We concluded that HVD was probably related to the external circumferential diaphragm that explained the shrinkage of the aortic diameter. Nevertheless, it is interesting to notice the rapid development of this “pannus-like” structure, in less than a year (Additional files 1, 2, 3). Concerning its potential cause, several hypotheses are put forward. It may be related to the surgical techniques as the structure was near the pledgets used for our surgical interrupted suture, reinforced with biological fibrin glue.

Fig. 2
figure 2

Transthoracic echocardiography, long axis parasternal view with same patient early post-operatory (A) and (C) and 1-year follow up exam (B) and (D), showing a shrinkage of the aortic annulus at 1 year (measured at the level of the yellow arrows), from 18 to 14 mm (A) and (C) together with a transaortic mean gradient and maximal velocity by Doppler continuous wave interrogation, from 2,4 m/sec and 12 mmHg (B) to 4,6 m/sec and 59 mmHg (D)

As for the surgical implantation technique, we used either a multiple, single, interrupted, non-everting, pledged reinforced U-stiches to implant the composite graft in supra-annular position (n = 33, 36%) or everting stitches to implant the prosthesis in intra-annular position (n = 58, 64%). After univariate analysis, we found no correlation between the technique and the event of bioprosthetic valve dysfunction (Table 2, p = 0.47).

We can also consider the role of the immune system. The glutaraldehyde bioconduit treatment is meant to eliminate immunogenic proteins [14]. Nevertheless, multiple observational studies identified xenograft porcine components (extracellular matrix specific glycans, galactose-a-1,3-galactose and N-glycolylneuraminic acid) who trigger host antibody formation [15,16,17,18,19]. Sustaining the same immunogenic hypothesis, a group from Munich studied the effects of a pericardial porcine, No-React® patch and they found sterile abscess formation that was suspected to be an immunogenic reaction, a xenogeneic complement-mediated graft rejection [20]. The question of patient-prosthesis mismatch can be raised because the smallest sizes of prostheses seemed to be more sensitive to hemodynamic deterioration. We believe that this observation is not related to an initial mismatch but to the fact that a decrease in EOA logically has more impact on a small prosthesis than on a larger one.

In light of all of the above, we speculate an early inflammatory reaction could have been triggered by the biological conduit itself, favored by the use of biological glue or/and by the surgical technique (all patients were implanted using a multiple, single, interrupted, everting or non-everting suture with pledged reinforced U-stiches). Lastly, concerning the surgical technique, we note that other teams (Carrel et al., Stefanelli et al., Sahin et al., Kaya et al., Galinanes et al.) reported using as technique of conduit implantation either a running suture or an interrupted, nevertheless, they were not confronted with this problem [13, 21,22,23,24].

Six patients were diagnosed with graft endocarditis. Diagnosis was made according to the modified Duke criteria and was often difficult, due to a combination of atypical clinical, biological, radiological and echocardiographic observations [25]. These features are not specific to this conduit and are usual after Bentall surgery. These results are surprising insofar as one of the advantages put forward in favor of these conduits is their expected low rate of reinfection.

In this regard, Galinanes et al. reported excellent late results (1 case of conduit infective endocarditis in 10 years) [13]. Siniawski et al., Musci et al. and Wendt et al. suggested the same reassuring results [26, 27, 31]. Recently, Stefanelli et al. also reported satisfactory results, with freedom from the BioConduit® graft infection of 95.7% at 5 and 15 years (CI 0.95) [22]. However, larger cohorts are still needed to confirm these results.

Previous studies reported a high number of aorto-ventricular disconnections especially related to the initial Shelhigh’s biological conduit [21, 23, 24, 28]. The Shelhigh graft was a stentless, valved, No-React® biological conduit implanted mostly in the early 2000s. Due to unexpected conduit disconnections, in 2007, the United States FDA published a notification, the product being retried of the market [28, 30]. Despite this drawback, in 2013, Musci et al. published reassuring results after 11 years of follow-up of 255 patients implanted with Shelhigh initial graft [30]. The Berlin group reported a low reinfection rate (0.78% early and 2.35% late reinfection) concluding that patients’ outcome was dependent of their surgical urgency. Afterwards, a new version of the graft, the BioIntegral Surgical, BioConduit™ was released on the market. Regarding its predecessor (the Shelhigh conduit), while some teams report relatively satisfactory early and late follow-up results (Galinanes et al., Musci et al.), other teams report some dreadful complications (Sahin et al., Kaya et al., Carrel et al., Reineke et al., Sadeque et al.) referring to the high rate of endocarditis and of atrio-ventricular disconnections with proximal false aneurysm formation [13, 21, 23, 24, 27,28,29,30, 32]. In contrast, they report conduit hemodynamic dysfunction only in a few cases. Reineke et al. describes 8 cases (2.3%) with structural valve deterioration at late follow up. In contrast, Kaya et al. reports 3 cases (1.7%) with hemodynamic failure at early follow-up, without detailing. Concerning the BioConduit™, Stefanelli et at publishes satisfactory early and long-term results [22]. In our series, we report 10 cases (13%) presenting criteria for non-structural hemodynamic valve dysfunction at 1 year and 4 of them fulfilling criteria for failure, needing reintervention. This was an unexpected finding, as to our knowledge, such a high incidence of biological conduit hemodynamic dysfunction was never cited before.

There are several limitations of this study. This was a prospective but single-center, non-randomized study with lack of group control. Even though there were 91 patients included, the population analyzed was rather small and heterogenous, gathering elective as well as emergent cases. Being a single institution study, with 4 surgeons and 2 techniques, while representing a limitation, could contribute to reducing biases related to the use of multiple techniques.

Conclusions

This study reports the early and midterm results of the newest all-biological valved conduit. Even though it is designed to achieve superior hemodynamics by excluding the valvular stent, our study reveals an abnormally high rate of early prosthetic non-structural dysfunction and failure that appears to be mostly related to a multifactorial progressive shrinkage of the aortic annulus. Regarding graft infections, we observed that endocarditis is not rare, despite conduits all-biological structure. Lastly, there are still some cases of conduit proximal anastomotic detachments.

Availability of data and materials

The datasets and the materials used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

BVD:

Bioprosthetic valve dysfunction

BVF:

Bioprosthetic valve failure

EOA:

Effective orifice area

HVD:

Hemodynamic valve deterioration

NSVD:

Nonstructural valve deterioration

PPM:

Patient-prosthesis mismatch

SVD:

Structural valve deterioration

TAVR:

Transcatheter aortic valve replacement

TEE:

Transesophageal echocardiogram

TTE:

Transthoracic echocardiogram

References

  1. Etz CD, Homann TM, Rane N, Bodian CA, Di Luozzo G, Plestis KA, et al. Aortic root reconstruction with a bioprosthetic valve conduit; a consecutive series of 275 procedures. J Thorac Cardiovasc Surg. 2007;133:1455–63.

    Article  Google Scholar 

  2. Bentall H, DeBono A. A technique for complete replacement of the ascending aorta. Thorax. 1968;23:338–9.

    Article  CAS  Google Scholar 

  3. Kouchoukos NT, Wareing TH, Murphy SF, Perrillo JB. Sixteen‐year experience with aortic root replacement. Ann Surg. 1991;214: 308–18. Results of 172 operations.

  4. Cherry C, DeBord S, Hickey C. The modified Bentall procedure for aortic root replacement. AORN J. 2006;84(1):52–5, 58–70; quiz 71–4.

  5. Abolhoda A, Yu S, Oyarzun JR, McCormick JR, Bogden JD, Gabbay S. Calcification of bovine pericardium: glutaraldehyde versus No-React biomodification. Ann Thorac Surg. 1996;62(1):169–74.

    Article  CAS  Google Scholar 

  6. Abolhoda A, Yu S, Oyarzun JR, Allen KR, McCormick JR, Han S, Kemp FW, Bogden JD, Lu Q, Gabbay S. No-react detoxification process: a superior anticalcification method for bioprosthesis. Ann Thorac Surg. 1996;62(6):1724–30.

    Article  CAS  Google Scholar 

  7. BioIntegral Surgical. BioConduit. http://www.biointegral-surgical.com/. Accessed 28 May 2021

  8. Akins CW, Craig Miller D, Turina EI, et al. Guidelines for reporting mortality and morbidity after cardiac valve interventions. Ann Thorac Surg. 2008;85:1490–5.

    Article  Google Scholar 

  9. Généraux P, Piazza N, Alu M et all. Valve Academic Research Consortium 3: updated endpoint definitions for aortic valve clinical research. JACC. 2021:2717–46

  10. Lancellotti P, Pibarot P, et al. 2016 ESC Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging endorsed by the Chinese Society of Echocardiography, the Inter-American Society of Echocardiography, and the Brazilian Department of Cardiovascular Imaging. Eur Heart J. 2016;17:589–90.

    Google Scholar 

  11. Lancellotti P, Pibarot P, Chambers J, Edvardsen T, Delgado V, Dulgheru R, Pepi M, et al. Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging endorsed by the Chinese Society of Echocardiography, the Inter-American Society of Echocardiography, and the Brazilian Department of Cardiovascular Imaging. Eur Heart J Cardiovasc Imaging. 2016;17(6):589–90. https://doi.org/10.1093/ehjci/jew025.

    Article  Google Scholar 

  12. Stefanelli G, Pirro F, Machionne A, Bellisario A, Weltert L. Long-term follow up after Bentall operation using a stentless Shelhigh NR-2000 bio-conduit. J Card Surg. 2020:1–8.

  13. Galinanes M, Meduoye A, Ferreira I, Sosnowski A. Totally biological composite aortic stentless valved conduit for aortic root replacement: 10-year experience. J Cardiothorac Surg. 2011;6:86.

    Article  Google Scholar 

  14. Kim KM, Herrera GA, Battarbee HD. Role of glutaraldehyde in calcification of porcine aortic valve fibroblasts. Am J Pathol. 1999;154:843–52.

    Article  CAS  Google Scholar 

  15. Lovekamp JJ, Simionescu DT, Mercuri JJ, Zubiate B, Sacks MS, Vyavahare NR. Stability and function of glycosaminoglycans in porcine bioprosthetic heart valves. Biomaterials. 2006;27:1507–18.

    Article  CAS  Google Scholar 

  16. Perota A, Lagutina I, Duchi R, Zanfrini E, Lazzari G, Judor JP, Conchon S, Bach JM, Bottio T, Gerosa G, et al. Generation of cattle knockout for galactose-α1,3-galactose and N-glycolylneuraminic acid antigens. Xenotransplantation. 2019;26: e12524.

    Article  Google Scholar 

  17. Lee W, Long C, Ramsoondar J, Ayares D, Cooper DK, Manji RA, Hara H. Human antibody recognition of xenogeneic antigens (NeuGc and Gal) on porcine heart valves: could genetically modified pig heart valves reduce structural valve deterioration? Xenotransplantation. 2016;23:370–80.

    Article  Google Scholar 

  18. Hawkins RB, Frischtak HL, Kron IL, Ghanta RK. Premature bioprosthetic aortic valve degeneration associated with allergy to galactose-alpha-1,3-galactose. J Card Surg. 2016;31:446–8.

    Article  Google Scholar 

  19. Jin R, Greenwald A, Peterson MD, Waddell TK. Human monocytes recognize porcine endothelium via the interaction of galectin 3 and alpha-GAL. J Immunol. 2006;177:1289–95.

    Article  CAS  Google Scholar 

  20. Elmer C. Experimental work-up of the bovine NO-REACT pericardial patch used in cardiac surgery in conjunction with late complications after its application. Doctoral Thesis. Institute of Surgical Research, Ludwig-Maximilians-University-Munich, Germany, 2007.

  21. Carrel TP, Schoenhoff FS, Schmidli J, Stalder M, Eckstein FS, Englberger L. Deleterious outcome of no-react-treated stentless valved conduits after aortic root replacement: Why were warnings ignored? J Thorac Cardiovasc Surg. 2008;136:52–7.

    Article  Google Scholar 

  22. Stefanelli G, Pirro F, Machionne A, Bellisario A, Weltert L. Long-term follow up after Bentall operation using a stentless Shelhigh NR-2000 bio-conduit. J Card Surg. 2020;35:1–8.

    Article  Google Scholar 

  23. Sahin A, Müggler O, Sromicki J, et al. Long-term follow-up after aortic root replacement with the Shelhigh biological valved conduit: a word of caution! Eur J Cardiothorac Surg. 2016;10:1093–178.

    Google Scholar 

  24. Kaya A, Heihmen RH, Kelder JC, Schepens Marc A, et al. Stentless biological valved conduit for artic root replacement: Initial experience with the Shelhigh BioConduit model NR 200C. J Thorac Cardiovasc Surg. 2011;141:1157–62.

    Article  Google Scholar 

  25. Habib G, Lancellotti P, Antunes MJ, Bongiorni MG, Casalta JP, Del Zotti F, Dulgheru R, El Khoury G, Erba PA, Iung B, Miro JM, Mulder BJ, Plonska-Gosciniak E, Price S, Roos-Hesselink J, Snygg-Martin U, Thuny F, Tornos Mas P, Vilacosta I, Zamorano JL; ESC Scientific Document Group. 2015 ESC Guidelines for the management of infective endocarditis: The Task Force for the Management of Infective Endocarditis of the European Society of Cardiology (ESC). Endorsed by: European Association for Cardio-Thoracic Surgery (EACTS), the European Association of Nuclear Medicine (EANM). Eur Heart J. 2015 Nov 21;36(44):3075–3128.

  26. Siniawski H, Lehmkuhl H, Weng Y, Pasic M, Yankah C, Hoffmann M, Behnke I, Hetzer R. Stentless aortic valves as an alternative to homografts for valve replacement in active infective endocarditis complicated by ring abscess. Ann Thorac Surg. 2003;75(3):803–8; discussion 808.

  27. Musci M, Siniawski H, Knosalla C, Grauhan O, Weng Y, Pasic M, Meyer R, Hetzer R. Early and mid-term results of the Shelhigh stentless bioprosthesis in patients with active infective endocarditis. Clin Res Cardiol. 2006;95(5):247–53.

    Article  Google Scholar 

  28. Reineke DC, Kaya A, Heinisch PP, et al. Long-term follow-up after implantation of the Shelhigh® No-React® complete biological aortic valved conduit. Eur J Cardio-Thorac Surg. 2016;50:98–104.

    Article  Google Scholar 

  29. Gabbay, Letter to the Editor, JTCVS, Vol. 137, Number 5, 1293–94

  30. Musci M, Siniawski H, Pasic M, et al. Surgical therapy in patients with active infective endocarditis: seven-year single centre experience in a subgroup of 255 patients treated with the Shelhigh® stentless bioprosthesis. Eur J Cardio Thorac Surg. 2008;34:410–7.

    Article  Google Scholar 

  31. Wendt D, Raweh A, Knipp S, El Gabry M, et al. Comparison of mid-term haemodynamic performance between the BioValsalva and the BioIntegral valved conduits after aortic root replacement. Interactive Cardio Vasc Thorac Surg Adv. 2016;23:1–6.

    Google Scholar 

  32. Sadeque SAA, Salhiyyah K, Livesey S. Pseudoaneurysm formation in the BioIntegral no-react prosthesis: does it share the problems of its predecessor? Eur J Cardiothorac Surg. 2017;52(2):397.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This research has received no external funding so there is no funding body to acknowledge.

Author information

Authors and Affiliations

Authors

Contributions

The corresponding author (RB) and YL-B conceived and designed the analysis. AG collected data and conceived the tables. YL-B performed the statistical analysis. All authors discussed the results and contributed to the final manuscript. The corresponding author and JP wrote the paper. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Roxana Botea.

Ethics declarations

Ethics approval and consent to participate

The prospective, observational study was performed in accordance with the principles outlined in the Declaration of Helsinki and it was approved by the Institutional Review Board of the Rangueil University Hospital of Toulouse, France (number RnlPH 2022-50). Informed consent to participate in the study was obtained from participants. Patients were informed that their codified data would be used for the study.

Consent for publication

Participants have given written consent to publishing research findings.

Competing interests

None of the authors have any conflict of interest to declare.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1. Video 1 Transthoracic echocardiography, long axis parasternal view. Exam at 1-year follow up exam showing a diminished and hyperechogenic, with a pannus-like structure at the level of the aortic annulus.

Additional file 2. Video 2 Transthoracic echocardiography, long axis parasternal view. Exam at 1-year follow up exam showing a diminished and hyperechogenic, with a pannus-like structure at the level of the aortic annulus.

Additional file 3. Video 3 Transthoracic echocardiography, long axis parasternal view. Exam at 1-year follow up exam showing a diminished and hyperechogenic, with a pannus-like structure at the level of the aortic annulus.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Botea, R., Lavie-Badie, Y., Goicea, A. et al. Early and midterm outcomes of a bentall operation using an all-biological valved BioConduit™. J Cardiothorac Surg 17, 325 (2022). https://doi.org/10.1186/s13019-022-02073-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-022-02073-5