Thromboembolic stroke in patients with a HeartMate-II left ventricular assist device – the role of anticoagulation

Abstract

Background and purpose

It is unknown what the optimal anticoagulant level is to prevent thromboembolic stroke in patients with left ventricular assist device (LVAD) support. We aimed to evaluate the relation between coagulation status and the occurrence of thromboembolic stroke in HeartMate-II LVAD assisted patients.

Methods

Thirty-eight consecutive patients with a HeartMate-II LVAD were included. Coagulation status was classified according to INR and aPTT ratio at: 1) the moment of first thromboembolic stroke; and 2) during the two weeks preceding the first thromboembolic stroke to assess long-term coagulation status. In patients without stroke, coagulation status was determined just before heart transplant, VAD explantation or death, whichever came first, and at two weeks preceding these surrogate endpoints. Based on coagulation status, patients were divided in two groups: Group I (reference group) was defined as INR below 2 and aPTT ratio below 1.5; Group II (adequate anticoagulation) as INR above 2 or aPTT ratio above 1.5. Logistic regression analysis was performed to assess the odds ratio for developing stroke for patients with adequate anticoagulation compared to the reference Group.

Results

Thromboembolic stroke occurred in six (16 %) patients, none within 2 weeks after LVAD implantation. Considering coagulation status at the time of event, patients in coagulation Group II had no decreased risk for thromboembolic stroke (OR 0.78; 95 % CI 0.12–5.0). Results for coagulation status 2 weeks prior of event could not be calculated as all six strokes occurred in Group II.

Conclusion

In our experience anticoagulation within predefined targets is not associated with a reduced thromboembolic stroke risk in patients with a HeartMate-II LVAD on antiplatelet therapy. However, no firm statement about the effect of either anticoagulant or antiaggregant therapy can be made based on our study. A larger randomized study is needed to support the hypothesis that there may be no additional benefit of coumarin or heparin therapy compared with antiplatelet therapy alone.

Peer Review reports

Background

Left ventricular assist devices (LVADs) have become an increasingly important therapeutic option in the treatment of end-stage heart failure. LVADs were initially introduced as ‘bridge-to-transplantation’. However, there is limited availability of donor organs and many patients with end-stage heart failure are not eligible for cardiac transplantation. Therefore, LVADs are applied as destination therapy as well, especially as outcomes for these patients have been demonstrated to be favorable when compared to medical therapy [14]. This results in more LVAD implantations, but also in a longer duration of implantation, which may increase the risk of complications including infections and stroke.

The reported incidence of stroke during LVAD support varies from 2 % to 47 % in studies evaluating different types of LVADs [58]. Anticoagulation is widely used for the prevention of thromboembolic stroke in LVAD patients, but the optimal therapeutic regime has been a matter of debate and no generally accepted guideline currently exists. Prophylactic administration of anticoagulants may enhance the risk of systemic bleeding and hemorrhagic stroke, which is already substantial due to LVAD-induced thrombopenia and acquired von Willebrand syndrome [9, 10]. It is therefore difficult to determine which anticoagulation regime is adequate.

The purpose of the current study was to evaluate the association between anticoagulation status and the occurrence of thromboembolic stroke in a single-centre cohort of patients treated with the HeartMate-II LVAD.

Methods

Study population and data sources

The study population consisted of all consecutive patients who underwent LVAD implantation in the Utrecht University Medical Center from 2006 until 2009. Only patients treated with the currently most frequently used continuous flow pump HeartMate-II LVAD were included in the analyses.

Demographic data, as well as data on the etiology of heart disease, the presence of risk factors for atherosclerosis, preoperative laboratory measurements and treatment characteristics were obtained through chart review. Long term outcome data were obtained through review of outpatient records. Data on the occurrence of cerebrovascular events were obtained through retrospective re-evaluation of imaging studies, in addition to review of clinical records and radiology reports.

In all cases, the intention of treatment with an LVAD was bridge-to-transplantation, therefore only patients who were eligible for cardiac transplant had LVAD implantation. All patients had a New York Heart Association (NYHA) classification IIIb or IV and received maximum medical treatment, including intravenous inotropic support. The decision for type of assist device depended on device availability and surgeon’s preference rather than on patient characteristics.

Anticoagulation protocol

Patients in whom a HeartMate type-II was implanted received acetylsalicylic acid 100 mg daily and were additionally treated with heparin as soon as possible with the aim to keep the activated partial thromboplastin time (aPTT) ratio between 1.5–2.0, which mostly resulted in an aPTT of 50–70 s. When discharge from the ICU was within reach, usually between 1 and 2 weeks postoperatively, acenocoumarol was started with a target International Normalized Ratio (INR) for outpatients of 2.0–3.0. Heparin was discontinued once an INR of 2.5 had been reached. Acetylsalicylic acid 100 mg daily was continued in all cases. Low molecular weight heparin was not used during the study period.

Outcomes and definitions

The primary outcome for the current analysis was the occurrence of thromboembolic stroke, including transient ischemic attack (TIA) or lacunar stroke. All cerebrovascular events were adjudicated by three authors (AvD, AS, WvdB) based on clinical information (occurrence, type of impairment, reversibility) and radiological investigations (area, size and aspect of infarction). In case of multiple events in one patient, only the first event was analyzed.

Coagulation status

In patients with stroke, coagulation status was determined at the time of stroke and during the two weeks preceding the event. The latter analysis was considered important as the onset of clot development may well be begin days before the occurrence of a symptomatic embolus. In patients without stroke, coagulation status was determined just before heart transplant, VAD explantation or death, whichever came first, and at two weeks preceding these surrogate endpoints. In patients with on-going LVAD-support at the time of follow-up, coagulation status was based on the most recently recorded values and values recorded during the two weeks before.

Patients were divided in two groups based on a combined assessment of the INR and aPTT ratio. Group I (reference group) included patients with both INR and aPTT ratio below the target of 2.0 and 1.5, respectively. Group II (adequate anticoagulation) included patients with either INR above 2.0 or aPTT ratio above 1.5. Patients who did not receive heparin had no mandatory aPTT measurements. Hence, in case of missing aPTT values they apparently were not on heparin therapy and were attributed an aPTT ratio of 1.

Statistical analysis

Continuous data were presented as median with interquartile range. Categorical data were presented as percentages. Differences between patients with different coagulation status were tested with the Fisher’s Exact test for categorical variables, and with the Mann-Whitney U-test for continuous variables that were not normally distributed. We imputed missing coagulation parameters using single imputation with the linear regression method with addition of a random error term. In 18 % of patients one or more values were missing, although not necessarily within two weeks preceding an event. The mean frequency of coagulation measurements within the two week preceding the event was 4.4 (STD 4.6). The occurrence of stroke was presented as both cumulative incidences (i.e. the percentage of patients with stroke for each category), and incidence rate (i.e. number of events per patient-years). Univariate logistic regression analysis was performed to assess the odds ratio for developing stroke for the two different categories of coagulation status at the time of the event. Group I was used as the reference category. All baseline characteristics were considered potential confounders [Table 1]. The influence of potential confounders was investigated by conducting multiple bivariate analyses. Any confounder that influenced the odds ratio for developing stroke by >10 % in the bivariate analyses was included in the multivariate model. Similar analyses were performed for mean coagulation status during the two weeks preceding the event. Kaplan-Meier curves for the two categories of coagulation status at the event were constructed. A p value < 0.05 (two sided) was considered to be statistically significant.

Table 1 Baseline characteristics according to coagulation status: Group I (low anticoagulation status) included patients with both INR below 2.0 and aPTT ratio below 1.50; Group II (adequate anticoagulation status) included patients with either INR ≥ 2.0 or aPTT ratio ≥ 1.50

Ethics

The study complies with the Declaration of Helsinki and good clinical practice guidelines. We obtained ethics committee approval from the University Medical Center Utrecht. Written informed consent was waived by the ethics committee. Patient information was anonymized and de-identified prior to analysis.

Results

Patients

From 2006 through 2009 a total of 41 consecutive patients underwent a HeartMate-II LVAD implantation in our hospital. Three patients were excluded from the primary analysis on thromboembolic stroke; one suffered hemorrhagic stroke (aPTT ratio at that time was 2.60), one post-anoxic stroke, and one symptomatic MELAS (mitochondrial myopathy, encephalopathy, lactate acidosis and stroke like episodes). Tables 1 and 2 show the baseline characteristics and clinical course data for the remaining 38 patients. Over one-fifth of patients required preoperative mechanical circulatory support with an intra-aortic balloon pump or a temporary VAD (e.g. Impella percutaneous left ventricular assist device). Median duration of support was 279 days (IQR 134 – 540); median duration of follow-up was 275 days (IQR 75 – 522). No patients underwent concomitant RVAD-implantation. Complication rate for tamponade, rethoracothomy and gastro-intestinal bleeding was higher in Group II, but due to low numbers didn’t reach statistical significance. Half of these complications occurred in the first 48 h after implantation, and two-third occurred within the first week.

Table 2 Clinical course and outcome according to coagulation status: Group I (low anticoagulation status) included patients with both INR below 2.0 and aPTT ratio below 1.50; Group II (adequate anticoagulation status) included patients with either INR ≥ 2.0 or aPTT ratio ≥ 1.50

Coagulation status

At the time of the primary or surrogate event, the reference Group I consisted of 12 patients and Group II consisted of 26 patients. When calculating mean coagulation during the two weeks preceding the event, 8 patients fell into Group I, and Group II consisted of 30 patients. The duration of sub-therapeutic anticoagulation range from several hours to one week. No differences were demonstrated between these groups with regard to baseline characteristics, in particular in pre-LVAD right ventricular failure, which parameter was previous associated with stroke development [11].

Thromboembolic stroke

Overall, 6 thromboembolic strokes occurred in 6 patients (16 %). All occurred after 14 days after LVAD implantation or later. Median time of stroke occurrence was 43 days (IQR 14–440) after LVAD implantation. In general, more strokes occurred in Group II compared with the reference Group I [Table 2] [Fig. 1].

Fig. 1
figure 1

Thromboembolic stroke incidence according to coagulation status at event. Patients within Group I are positioned in the square in the lower left hand corner; Group II consists of the 3 squares that enclose Group I. O = no thromboembolic stroke. * = thromboembolic stroke

Fig. 2
figure 2

Thromboembolic stroke incidence according to coagulation status 2 weeks prior to event. Patients within Group I are positioned in the square in the lower left hand corner; Group II consists of the 3 squares that enclose Group I. O = no thromboembolic stroke.  = thromboembolic stroke

Considering the analysis for coagulation status at the time of stroke, the Kaplan-Meier estimates for 1-year stroke free survival were 100 % for patients in Group I and 84 % for patients in Group II [Fig. 2]. When coagulation status was based on the two weeks preceding the event the Kaplan-Meier estimates for 1-year stroke free survival were 100 % for patients in Group I and 86 % in Group II [Figs. 3 and 4].

Compared with reference Group I, the adjusted odds ratio for the development of thromboembolic stroke was 0.53 (95 % CI 0.06–4.6) for coagulation Group II [Table 3]. When coagulation status was based on the two weeks preceding the event, the odds ratio could not be calculated as all strokes occurred in patients with adequate anticoagulation (Group II).

Table 3 Influence of coagulation status on the occurrence of thromboembolic stroke – results of logistic regression analyses

Discussion

The main finding from our study is that patients assisted by a HeartMate-II LVAD with an INR or aPTT above predefined targets did not have a reduced stroke incidence. Another important finding was that there was a trend towards more early complications, e.g. rethoracothomy and gastrointestinal bleeding, in patients with adequate anticoagulation.

It is widely assumed that the optimal INR to minimize the risk of both thromboembolic and hemorrhagic events needs to be between 1.5 and 2.5, in addition to the use of aspirin. A study in out-patients showed that the risk of thrombotic events increased when the INR was below 1.5, whereas the risk of hemorrhagic events sharply increased with an INR above 2.5 [6]. One important limitation of that study is that INR levels were recorded only at monthly intervals and at the time of a clinical event, whereas INRs for outpatients can change widely and over much shorter time periods according to patient conditions. Furthermore, patients were not included in the study until they were successfully discharged. Thus, patients who had peri-operative bleeding or thromboembolic events before discharge that was not related to the degree of oral anticoagulation were excluded. Also, numbers were very low within this sub-study in out-patients only and no statistics were performed to show that the numbers were statistically significant [6]. Several other studies also found that suboptimal anticoagulation was associated with neurological events, but it all concerns observational studies with low absolute numbers and several other risk factors were identified [12, 13]. Another limitation is that a single measurement below the target value, often at the moment stroke occurred, was used for analysis. Moreover, hemorrhagic stroke was associated with increased coagulation status which is the reason for some to advocate a mild anticoagulation protocol [14].

A limitation of our study is its retrospective design, withholding an increased risk of missing values. To overcome this problem several measures were taken. Firstly, the INR was, when appropriate, calculated from the PT ratio and the aPTT ratio from the heparin ratio. However, although the mentioned tests are related, each marker is calculated using different types of reagent and is therefore not freely interchangeable. The results of these calculations must therefore be seen as estimates. Secondly, our hospital protocol commands aPTT (or heparin ratio) being measured 4 times daily when heparin is administered. As a consequence, if an aPTT or heparin ratio value was unknown, it was presumed that the patient had no heparin therapy at that time and an aPTT ratio of 1 was filled in. However, in outpatients on coumarin, INR measurements are performed with larger intervals and therefore a missing value at a certain moment does not exclude coumarin therapy. To overcome this, an imputation procedure was performed to fill in the remaining blanks. Data omission rarely occurs at random, but is often related to other observed patient characteristics. Excluding subjects with missing values not only leads to loss of statistical power, but also to biased results [1517]. To decrease bias and increase statistical efficiency, we therefore chose to impute missing values, rather than to perform complete-case analyses.

Another possible limitation is that although many risk factors were assessed, we didn’t systematically collect pre-implant atrial fibrillation or the CHA2DS2-Vasc score in our patients, which is associated with a higher risk of thromboembolic events in patients with continuous flow LVAD [18]. None of the 6 stroke patients had atrial fibrillation at the time of stroke, so the impact on our results is limited.

Throughout the study period various models of LVAD have been used and results of this study may not be translated to the other used devices. We chose to restrict the analysis to one particular LVAD type (i.e. Heart Mate II) to rule out that the observed differences in thromboembolic events were due to the different type of LVAD used.

Fig. 3
figure 3

Kaplan-Meier survival curves showing stroke free survival according to coagulation status at event (Fig. 3). ______ Group I ………. Group II

Fig. 4
figure 4

Kaplan-Meier survival curves showing stroke free survival according to coagulation status 2 weeks prior to event (Fig. 4). ______ Group I ………. Group II

The three patients with primary intracranial hemorrhages, global ischemia and MELAS, were left out of the case control analyses, because these events are not related to thromboembolism. The low incidence of intracranial hemorrhage, however, may indicate that anticoagulation with coumarin or heparin does not lead to an increased risk of intracranial hemorrhage. It is well known that dual or triple therapy increase the risk of bleeding complications [19]. This is further illustrated by the higher incidence of tamponade and rethoracothomy in the group of patients with adequate anticoagulation status.

To our knowledge, no randomized clinical trial on anticoagulation in patients with VAD, has ever been performed. We have previously performed a systematic review on the relation between anticoagulation protocols and occurrence of stroke after LVAD implantation, in which we found that platelet aggregation inhibitors are the most important in preventing thromboembolic stroke and should be part of the anticoagulation protocol [20]. The World’s Longest-Supported HeartWare Ventricular Assist Device Patient has been managed on low-dose aspirin (100 mg three times per week) without thrombotic complications [21].

Based on the results of our study it is not clear if postoperative heparin decrease the number of thromboembolic strokes in patients with a HeartMate-II support. Since heparin therapy may increase the risk of short-term systemic bleeding complications these findings indicate that the postoperative use of heparin after HeartMate II implantation may be refrained or postponed in selected patients with an anticipated increased bleeding risk.

Although not undoubtedly effective in preventing long term stroke in our study, the recently reported multicenter increase in the incidence of pump thrombosis with the HeartMate II LVAD may be an additional reason apart from patient characteristics to maintain anticoagulation with coumarins [22]. Both in-patient and device complication remains a challenge for bioengineering, which advances may be even more important than anticoagulants regimes as preclinical studies in LVAD technology demonstrates that design changes may offer a more robust alternative to antiplatelet drugs [23]. Within the study population, no LVADs needed to be replaced, so probably there was no occurrence of (significant) pump thrombosis, but unfortunately we didn’t systematically assessed this endpoint.

The suggestion from the results of our study that antiplatelet therapy could be sufficient without the addition of coumarins should be supported by larger studies, and more important by a qualitative assessment of the coagulation status such as thromboelastometry and platelet function assays besides the classical parameters INR, PTT, and platelet count. With such information, it would be possible to calibrate the antithrombotic therapy, deciding dose variation or withdrawal of either anticoagulant or antiaggregant therapy, on a mechanistic basis.

Conclusion

Thromboembolic stroke remains a feared complication in patients assisted by a HeartMate-II LVAD. A vast majority of strokes occurred despite adequate anticoagulation according to the protocol of our institute. Methodological shortcomings of our study prevent a firm statement on the effect of coumarin or heparin in addition to antiplatelet therapy. It is currently unclear how stroke can be prevented in this patient category. A randomized clinical trial is warranted in order to improve outcome of this frail and growing patient category.

Abbreviations

aPTT:

Activated partial thromboplastin time

INR:

International Normalized Ratio

LVAD:

Left ventricular assist device

OR:

Odds ratio

PTT:

Partial thromboplastin time

MELAS:

Mitochondrial myopathy, encephalopathy, lactate acidosis and stroke like episodes

NYHA:

New York Heart Association

RVAD:

Right ventricular assist device

TIA:

Transient ischemic attack

VAD:

Ventricular assist device

References

  1. Frazier OH, Rose EA, Oz MC, Dembitsky W, McCarthy P, Radovancevic B, et al. Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantation. J Thorac Cardiovasc Surg. 2001;122(6):1186–95.

    Article  CAS  PubMed  Google Scholar 

  2. Kirklin JK, Naftel DC, Kormos RL, Stevenson LW, Pagani FD, Miller MA, et al. Second INTERMACS annual report: more than 1,000 primary left ventricular assist device implants. J Heart Lung Transplant. 2010;29(1):1–10.

    Article  PubMed  PubMed Central  Google Scholar 

  3. Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, et al. Long-term use of a left ventricular assist device for end-stage heart failure. N Engl J Med. 2001;345(20):1435–43.

    Article  CAS  PubMed  Google Scholar 

  4. Slaughter MS, Rogers JG, Milano CA, Russell SD, Conte JV, Feldman D, et al. Advanced heart failure treated with continuous-flow left ventricular assist device. N Engl J Med. 2009;361(23):2241–51.

    Article  CAS  PubMed  Google Scholar 

  5. John R, Kamdar F, Liao K, Colvin-Adams M, Miller L, Joyce L, et al. Low thromboembolic risk for patients with the Heartmate II left ventricular assist device. J Thorac Cardiovasc Surg. 2008;136(5):1318–23.

    Article  PubMed  Google Scholar 

  6. Boyle AJ, Russell SD, Teuteberg JJ, Slaughter MS, Moazami N, Pagani FD, et al. Low thromboembolism and pump thrombosis with the HeartMate II left ventricular assist device: analysis of outpatient anti-coagulation. J Heart Lung Transplant. 2009;28(9):881–7.

    Article  PubMed  Google Scholar 

  7. Schmid C, Weyand M, Nabavi DG, Hammel D, Deng MC, Ringelstein EB, et al. Cerebral and systemic embolization during left ventricular support with the Novacor N100 device. Ann Thorac Surg. 1998;65(6):1703–10.

    Article  CAS  PubMed  Google Scholar 

  8. Slater JP, Rose EA, Levin HR, Frazier OH, Roberts JK, Weinberg AD, et al. Low thromboembolic risk without anticoagulation using advanced-design left ventricular assist devices. Ann Thorac Surg. 1996;62(5):1321–7.

    Article  CAS  PubMed  Google Scholar 

  9. Crow S, Chen D, Milano C, Thomas W, Joyce L, Piacentino III V, et al. Acquired von Willebrand syndrome in continuous-flow ventricular assist device recipients. Ann Thorac Surg. 2010;90(4):1263–9.

    Article  PubMed  Google Scholar 

  10. Uriel N, Pak SW, Jorde UP, Jude B, Susen S, Vincentelli A, et al. Acquired von Willebrand syndrome after continuous-flow mechanical device support contributes to a high prevalence of bleeding during long-term support and at the time of transplantation. J Am Coll Cardiol. 2010;56(15):1207–13.

    Article  PubMed  Google Scholar 

  11. Nakajima I, Kato TS, Komamura K, Takahashi A, Oda N, Sasaoka T, et al. Pre- and post-operative risk factors associated with cerebrovascular accidents in patients supported by left ventricular assist device. -Single center's experience in Japan-. Circ J. 2011;75(5):1138–46.

    Article  PubMed  Google Scholar 

  12. Whitson BA, Eckman P, Kamdar F, Lacey A, Shumway SJ, Liao KK, et al. Hemolysis, pump thrombus, and neurologic events in continuous-flow left ventricular assist device recipients. Ann Thorac Surg. 2014;97(6):2097–103.

    Article  PubMed  Google Scholar 

  13. Morgan JA, Brewer RJ, Nemeh HW, Gerlach B, Lanfear DE, Williams CT, et al. Stroke while on long-term left ventricular assist device support: incidence, outcome, and predictors. ASAIO J. 2014;60(3):284–9.

    Article  PubMed  Google Scholar 

  14. Menon AK, Götzenich A, Sassmannshausen H, Haushofer M, Autschbach R, Spillner JW. Low stroke rate and few thrombo-embolic events after HeartMate II implantation under mild anticoagulation. Eur J Cardiothorac Surg. 2012;42(2):319–23.

    Article  PubMed  Google Scholar 

  15. Greenland S, Finkle WD. A critical look at methods for handling missing covariates in epidemiologic regression analyses. Am J Epidemiol. 1995;142(12):1255–64.

    CAS  PubMed  Google Scholar 

  16. Harrell Jr FE, Lee KL, Mark DB. Multivariable prognostic models: issues in developing models, evaluating assumptions and adequacy, and measuring and reducing errors. Stat Med. 1996;15(4):361–87.

    Article  PubMed  Google Scholar 

  17. Donders AR, van der Heijden GJ, Stijnen T, Moons KG. Review: a gentle introduction to imputation of missing values. J Clin Epidemiol. 2006;59(10):1087–91.

    Article  PubMed  Google Scholar 

  18. Koene RJ, Win S, Naksuk N, Adatya SN, Rosenbaum AN, John R, et al. HAS-BLED and CHA2DS2-VASc scores as predictors of bleeding and thrombotic risk after continuous-flow ventricular assist device implantation. J Card Fail. 2014;20(11):800–7.

    Article  PubMed  Google Scholar 

  19. Sorensen R, Hansen ML, Abildstrom SZ, Hvelplund A, Andersson C, Jorgensen C, et al. Risk of bleeding in patients with acute myocardial infarction treated with different combinations of aspirin, clopidogrel, and vitamin K antagonists in Denmark: a retrospective analysis of nationwide registry data. Lancet. 2009;374(9706):1967–74.

    Article  CAS  PubMed  Google Scholar 

  20. Backes D, van den Bergh WM, van Duijn AL, Lahpor JR, van Dijk D, Slooter AJ. Cerebrovascular complications of left ventricular assist devices. Eur J Cardiothorac Surg. 2012;42(4):612–20.

    Article  PubMed  Google Scholar 

  21. Critoph C, Chih S, Baumwol J. The World's Longest-Supported HeartWare Ventricular Assist Device Patient: 6 Years and Counting on Minimal Anticoagulation. Artif Organs. 2014;38(5):434–5.

    Article  PubMed  Google Scholar 

  22. Starling RC, Moazami N, Silvestry SC, Ewald G, Rogers JG, Milano CA, et al. Unexpected abrupt increase in left ventricular assist device thrombosis. N Engl J Med. 2014;370(1):33–40.

    Article  CAS  PubMed  Google Scholar 

  23. Bluestein D, Girdhar G, Einav S, Slepian MJ. Device thrombogenicity emulation: a novel methodology for optimizing the thromboresistance of cardiovascular devices. J Biomech. 2013;46(2):338–44.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Walter M. van den Bergh.

Additional information

Competing interests

Jaap R. Lahpor is consultant and member of the European Advisory Board Thoratec Europe Ltd. The other authors declare that they have no competing interests.

Authors’ contribution

All authors participated in design of the study, collection of the data, analysis of the data and critically reviewed the paper. All authors read and approved the final manuscript.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

van den Bergh, W.M., Lansink-Hartgring, A.O., van Duijn, A.L. et al. Thromboembolic stroke in patients with a HeartMate-II left ventricular assist device – the role of anticoagulation. J Cardiothorac Surg 10, 128 (2015). https://doi.org/10.1186/s13019-015-0333-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13019-015-0333-7

Keywords